Philips Medical Systems: Automated External Defibrillator(AED) of brand name "Philips HeartStart HS1 OnSite" (Model number M5066A). Device is designed for use on a person in sudden cardiac arrest who is unresponsive when shaken and not breathing normally

Recall number
Z-1916-2010
Recalling firm
Philips Medical Systems
Product
Automated External Defibrillator(AED) of brand name "Philips HeartStart HS1 OnSite" (Model number M5066A). Device is designed for use on a person in sudden cardiac arrest who is unresponsive when shaken and not breathing normally
Status
Terminated
Initiated
2010-04-30
Posted
2010-06-28
Terminated
2012-12-10
Reason
Seventeen AEDs failed production line testing (Final Acceptance Test).
Root cause
Nonconforming Material/Component
Action
The firm notified U.S. distributors on 4/30/2010 via phone call and e-mail to obtain consignee location information. A website providing information was activated on 4/30/2010. On 5/3/2010 Urgent Medical Device Recall letters, dated April 29, 2010 were sent to direct customers in the US. The letters explained the reason for the recall and Customers were asked to verify that they have the product subject to recall. Replacement AEDs will be provided at no charge to the customer. Field Safety Notices were attached to the letters and provided a list of actions for customers to take. Customers should check the serial number on their devices against the list provided, contact Philips Healthcare to arrange for replacement, and ensure that necessary staff are aware of the recall. For further information and support, Philips Healthcare should be contacted at 1-800-263-3342.
Quantity
205
Distribution
Worldwide Distribution -- USA, Canada, Japan, United Arab Emirates, Austria, Switzerland, Germany, Spain, France, Italy, and Norway.
Lot, serial or model codes
SERIAL NUMBERS--MODEL M5066A:  A10C-04048, A10C-04050, A10C-04300, A10C-04301, A10C-04302, A10C-04315, A10C-04316, A10C-04317, A10C-04318, A10C-04319, A10C-04320, A10C-04321, A10C-04322, A10C-04324, A10C-04325, A10C-04326, A10C-04327, A10C-04328, A10C-04330, A10C-04332, A10C-04333, A10C-04334, A10C-04335, A10C-04337, A10C-04338, A10C-04339, A10C-04340, A10C-04351, A10C-04352, A10C-04353, A10C-04354, A10C-04356, A10C-04357, A10C-04358, A10C-04359, A10C-04360, A10C-04361, A10C-04363, A10C-04364, A10C-04365, A10C-04366, A10C-04367, A10C-04368, A10C-04369, A10C-04611, A10C-04763, A10C-04812, A10C-04824, A10C-04824, A10C-05077, A10C-05078, A10C-05079, A10C-05080, A10C-05081, A10C-05083, A10C-05086, A10C-05087, A10C-05088, A10C-05089, A10C-05098, A10C-05151, A10C-05185, A10C-05191, A10C-05206, A10C-05207, A10C-05211, A10C-05212, A10C-05215, A10C-05216, A10C-05219, A10C-05233, A10C-05314, A10C-05315, A10C-05324, A10C-05397, A10C-05398, A10C-05429, A10C-05554, A10C-05569, A10C-05570, A10C-05571, A10C-05575, A10C-05578, A10C-05579, A10C-05580, A10C-05581, A10C-05583, A10C-05584, A10C-05585, A10C-05586, A10C-05587, A10C-05588, A10C-05591, A10C-05592, A10C-05593, A10C-05594, A10C-05595, A10C-05597, A10C-05599, A10C-05600, A10C-05751, A10C-05752, A10C-05759, A10C-05763, A10C-05764, A10C-05765, A10C-05767, A10C-05768, A10C-05769, A10C-05770, A10C-05771, A10C-05772, A10C-05775, A10C-05776, A10C-05777, A10C-05778, A10C-05779, A10C-05780, A10C-05781, A10C-05784, A10C-05785, A10C-05786, A10C-05787, A10C-05788, A10C-05789, A10C-05791, A10C-05792, A10C-05800, A10C-05913, A10C-05941, A10C-05968, A10C-05969, A10C-05970, A10C-05976, A10C-05977, A10C-05982, A10C-05983, A10C-06001, A10C-06002, A10C-06003, A10C-06006, A10C-06011, A10C-06013, A10C-06018, A10C-06020, A10C-06026, A10C-06027, A10C-06028, A10C-06354, A10C-06355, A10C-06356, A10C-06357, A10C-06358, A10C-06359, A10C-06360, A10C-06361, A10C-06364, A10C-06365, A10C-06366, A10C-06367, A10C-06368, A10C-06369, A10C-06370, A10C-06371, A10C-06372, A10C-06373, A10C-06374, A10C-06375, A10C-06376, A10C-06377, A10C-06378, A10C-06379, A10C-06381, A10C-06382, A10C-06383, A10C-06384, A10C-06385, A10C-06386, A10C-06387, A10C-06388, A10C-06390, A10C-06391, A10C-06392, A10C-06393, A10C-06394, A10C-06395, A10C-06396, A10C-06399, A10C-06400, A10C-06401, A10C-06402, A10C-06403, A10C-06404, A10C-06405, A10C-06406, A10C-06407, A10C-06408, A10C-06410, A10C-06411, A10C-06417, A10C-06418, A10C-06419, A10C-06420, A10C-06421, and A10C-06423.
510(k) numbers
["K020715"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1000524572
Firm city
Seattle
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register