FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: SOMATOM go.Up (Model #11061610) with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies.
- Recall number
- Z-1915-2019
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- SOMATOM go.Up (Model #11061610) with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies.
- Status
- Terminated
- Initiated
- 2019-05-28
- Posted
- not on file
- Terminated
- 2023-03-31
- Reason
- The potential sporadic performance problems may result in scanning workflow interruptions and unexpected user notifications. This may cause a delay in diagnosis or patient scans.
- Root cause
- Software design
- Action
- Urgent Medical Device Correction letters dated 5/24/19 were sent to customers.
- Quantity
- 9
- Distribution
- US Nationwide distributed in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, PA, SC, TX, VA, WA, WI, and Puerto Rico.
- Lot, serial or model codes
- Serial Number: 112552 112551 112555 112519 112542 112545 112541 112536 112573
- 510(k) numbers
- ["K173632"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28