Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics PAL Pelvic Alignment Level, Ref: PAL 400; Howmedica Osteonics Corp, 325 Corporate Drive, Mahwah, NJ 07430. Intended to determine proper anatomical orientation, selection, alignment and position of the acetabular and femoral components.

Recall number
Z-1915-2009
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics PAL Pelvic Alignment Level, Ref: PAL 400; Howmedica Osteonics Corp, 325 Corporate Drive, Mahwah, NJ 07430. Intended to determine proper anatomical orientation, selection, alignment and position of the acetabular and femoral components.
Status
Terminated
Initiated
2007-09-11
Posted
2009-08-26
Terminated
2009-10-15
Reason
The bubble in a vial of the Pelvic Alignment Level (PAL) was wider than the level marking. Upon opening the PAL and removing the vial, it was discovered that the vial was leaking
Root cause
Device Design
Action
Stryker Orthopaedics issued an "Important Product Removal" letter dated September 11, 2007 via Federal Express informing consignees of the affected devices. The firm requested that users identify and return any affected product and account for product by returning a Product Accountability Form via fax. For further information, contact Stryker Orthopaedics at 1-201-831-5825.
Quantity
615
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot Numbers: 969A01, 969A02, 969A03, 969A04, 969A05, 968B01, 968B02, 100126171-01 and samples.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IQO
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register