FDA Medical Device Recalls (openFDA)
Synthes (USA) Products LLC: DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43
- Recall number
- Z-1912-2016
- Recalling firm
- Synthes (USA) Products LLC
- Product
- DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43
- Status
- Terminated
- Initiated
- 2016-05-02
- Posted
- 2016-06-06
- Terminated
- 2016-12-13
- Reason
- Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could potentially be reproduced during use and reprocessing.
- Root cause
- Process design
- Action
- Depuy Synthes sent an Urgent Notice: Medical Device Recall letter on 05/02/2016 to all affected consignees. Consignees were instructed to do the following: review your inventory immediately to see if any of the affected products were being stored, and call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number and replacement order, to complete the Verification Section (page 3) and return with the product to Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132.
- Quantity
- 244 units
- Distribution
- US nationwide distribution.
- Lot, serial or model codes
- Catalog ID # 351.43 Lot numbers: XXX5071, 3945479, 4738380, XXX5072, 3997000, 4816738, XXX5073, 3996999, 4816739, A3CN449, 3997338, 4441056, 3007721, 4326050, 5407442, 3148892, 4307973, 5582819, 3175916, 4387338, 5722664, 3175202, 4441055, 5813837, 3175201, 4441079, 5786778, 3921562, 4559672, 5800844, 3928332, 4561383, 5903413, 3935789, 4619549, 5910995.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HTO
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28