Skip to the content

FDA Medical Device Recalls (openFDA)

Vital Signs Inc: Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color Zone: Green, Single Patient use, INTRAOSSEUS Module, Product # 7700GIO, Vital Signs, Totowa, NJ. The Broselow/Hinkle Pediatric Emergency System is a convenience kit that contains various modules/components used for the emergency treatment of pediatric patients. The modules are color coded to accommodate patients of various sizes.

Recall number
Z-1912-2009
Recalling firm
Vital Signs Inc
Product
Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color Zone: Green, Single Patient use, INTRAOSSEUS Module, Product # 7700GIO, Vital Signs, Totowa, NJ. The Broselow/Hinkle Pediatric Emergency System is a convenience kit that contains various modules/components used for the emergency treatment of pediatric patients. The modules are color coded to accommodate patients of various sizes.
Status
Terminated
Initiated
2009-06-30
Posted
2009-09-08
Terminated
2012-08-27
Reason
Not compatible: The luer connector on the Intraosseus needle provided does not fit securely on the luer connection of the IV Extension Set in the module.
Root cause
Device Design
Action
Vital Signs, Inc. issued an "Urgent Medical Device Correction" letter dated June 30, 2009 to all direct and indirect accounts by first class mail. Consignees were informed of the affected device, asked to complete a Confirmation form for return and to return unused product for replacement. For further questions, contact Vital Signs, Inc. at 1-973-790-1330 extension 404.
Quantity
23,429 kits
Distribution
Worldwide Distribution -- US, Canada, United Arab Emirate, Venezuela, Guam, Korea and United Kingdom.
Lot, serial or model codes
Model Numbers: 7700BIO, 7700GIO, 7700PIO, 7700RIO, 7700WIO, 7700YIO, 7700BIO2, 7700GIO2, 7700OIO2, 7700PIO2, 7700RIO2, 7700WIO2, 7700YIO2, 7730BLU5, 7730GRN5, 7730ORG5, 7730PUR5, 7730RED5, 7730WHI5, 7730YEL5, 7730FLY, 7730MOD, 7730ALS and 7730IALS; and Lot Numbers: (2008) 245F through 366F and (2009) 001K through 149K.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAA
Firm FEI
2242551
Firm city
Totowa
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register