Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

Recall number
Z-1911-2025
Recalling firm
Olympus Corporation of the Americas
Product
Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V
Status
Open, Classified
Initiated
2025-05-05
Posted
2025-06-05
Terminated
not on file
Reason
Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
Root cause
Labeling design
Action
Olympus notified consignees on 05/05/2025 via "URGENT Medical Device Correction" letter. Consignees were instructed to carefully review the notification, ensure all reprocessing and endoscopy unit personnel are completely knowledgeable and thoroughly trained on the updated reprocessing instructions, replace any copies of the reprocessing manual with the updated version (RC8415_02), complete and return the response form to recall web portal: https://olympusamerica.com/recall, enter the recall number: "0471" and complete the form as instructed, and forward the notification to affected customers if the affected product has been further distributed. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.
Quantity
12,195 units
Distribution
Worldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Guadeloupe, Guatemala, Honduras, Hong Kong, Iceland, India, Iraq, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Palestinian Authority, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Republic of the Congo, R¿union, Russia, Saudi Arabia, Senegal, Seychelles, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam, Yemen.
Lot, serial or model codes
UDI-DI: 04953170405563. Model Number: TJF-Q190V. All serial numbers.
510(k) numbers
["K220587"]
PMA numbers
not on file
Product code
NWB
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28