Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2330.

Recall number
Z-1909-2026
Recalling firm
Olympus Corporation of the Americas
Product
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2330.
Status
Open, Classified
Initiated
2026-03-25
Posted
2026-04-22
Terminated
not on file
Reason
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
Root cause
Device Design
Action
Olympus notified consignees on about 03/25/2026 via letter titled "URGENT: MEDICAL DEVICE REMOVAL." Consignees were notified of the issue and hazard involved and were instructed to examine inventory and OER-ELITE devices for the affected product. The current version has black lock levers. The previous versions have gray lock levers or integrated lock levers, which are the same color as the rest of the connector. Consignees were instructed that if they have sufficient quantities of the current version (black) to support reprocessing needs, stop using the previous version and set aside for return. They may proceed with using the current version, only. If they have insufficient quantities of the current version (black) available, and do not have another method of endoscope reprocessing readily available, they may continue using the previous version(s) consistent with the Instructions for Use, which require users to inspect the connecting tube prior to use. Transition to the current version as soon as possible by placing an order through Customer Service at 1-800-848-9024, option 2. When sufficient quantities of the current version (black) are obtained to support reprocessing needs, stop using the previous version and only use the current version going forward. Then arrange for return the previous version of the connector. Consignees were also requested to confirm receipt of the letter through the Olympus web portal and forward the notification to other users who may have the affected products if product were further distributed.
Quantity
1,829 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Hong Kong, Japan, South Korea and Taiwan.
Lot, serial or model codes
Model Number: MAJ-2330. UDI Number: 04953170404207. Lot Numbers: 09A, 0XA, 0YA, 0ZA, 11A, 12A, 24A, 25A, 26A, 27A, 28A, 29A, 2XA, 2YA, 2ZA, 31A, 32A, 33A, 34A, 35A, 36A, 38A, 39A, 3XA, 3YA, 3ZA, 41A, 42A, 43A, 44A, 45A, 46A, 47A, 48A, 49A, 4XA, 4YA, 4ZA, 51A.
510(k) numbers
["K201920"]
PMA numbers
not on file
Product code
FEB
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28