Skip to the content

FDA Medical Device Recalls (openFDA)

Breg Inc: Dressing Shldr, P/N 04908 Product Usage: These products are sterile dressings used as an insulation barrier between the Polar Care cold therapy pad and the skin. Sterile dressings are typically applied in a non-sterile environment after post-op and on occasion may be applied in a sterile environment. Breg has determined that a small percentage (approximately 0.22%) of these products may have been sent with unsealed pouches by the contract manufacturer. Customers will be contacted and instructed to inspect their inventory for any product with unsealed pouches and discard them.

Recall number
Z-1905-2017
Recalling firm
Breg Inc
Product
Dressing Shldr, P/N 04908 Product Usage: These products are sterile dressings used as an insulation barrier between the Polar Care cold therapy pad and the skin. Sterile dressings are typically applied in a non-sterile environment after post-op and on occasion may be applied in a sterile environment. Breg has determined that a small percentage (approximately 0.22%) of these products may have been sent with unsealed pouches by the contract manufacturer. Customers will be contacted and instructed to inspect their inventory for any product with unsealed pouches and discard them.
Status
Terminated
Initiated
2017-03-30
Posted
2017-05-02
Terminated
2017-06-23
Reason
They may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile.
Root cause
Employee error
Action
Breg sent an Urgent Medical Device Correction Letter to customers on 4/13/17. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: Please take the following actions: 1) Determine if any potentially affected products are in your stock by comparing each product's Lot and Date to the Lot Numbers and Expiration Dates in the table above. Refer to Picture 2 for the location of these fields on the product packaging. 2) Once you have matched Lot and Date, inspect the product to identify any open (unsealed) pouches. Refer to Picture 1 below. 3) Remove any open pouches from your inventory and discard. 4) Complete the Response Form included as Attachment A, and email or fax to Breg as follows: a. E-mail: bregfieldcorrection@breg.com: Subject: Sterile Dressing Correction b. Fax: 800-FAX-BREG (329-2734)- Subject: Sterile Dressing Correction Customers with questions are instructed to contact Breg Customer Care at (800) 321-0607.
Quantity
2,609 units total
Distribution
US Nationwide Distribution
Lot, serial or model codes
D163434
510(k) numbers
not on file
PMA numbers
not on file
Product code
KGX
Firm FEI
1000125378
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register