Mallinckrodt Inc: Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrodt, Inc. Cincinnati, OH 45237. Facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table.

Recall number
Z-1904-2010
Recalling firm
Mallinckrodt Inc
Product
Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrodt, Inc. Cincinnati, OH 45237. Facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table.
Status
Terminated
Initiated
2010-06-14
Posted
2010-06-25
Terminated
2012-02-25
Reason
Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative.
Root cause
Process control
Action
Covidien Issued an "Urgent Device Correction" notification dated June 14, 2010 informing their customers of the potential problem. Consignees were instructed that Field Service Engineers will be in contact arrange a service visit to perform corrections on the affected units. For further information, contact Covidien Product Monitoring at 1-800-778-7898.
Quantity
144 units
Distribution
Worldwide Distribution -- United States, Canada and China.
Lot, serial or model codes
Part Number: 750732, Serial Numbers: G28619, G28622, G28623, G28625, G28627, G28630, G28633, G28668, G28669, G28679, G28680, G28924, G28926, G28928, G28930, G28939, G28997, G29001, G29002, G29006, G29028, G29029, G29063, G29064, G29065, G29068, G29069, G29070, G29071, G29138, G29141, G29143, G29151, G29155, G29157, G29162, G29234, G29236, G29238, G29242, G29243, G29248, G29257, G29258, G29261, G29265, G29268, G29277, G29814, G30632, G30635, G30852, G30856, G31123, G31128, G31134, G31344, G31347, G31355, G31359, G31435, G31437, G31439, G31440, G31503, G31505, G31609, G31610, G31612, G31614, G31616, G32275, G32279, G32284, G32288, G32290, G32456, G32461, G32463, G32464, G32468, G32470, G32476, G32511, G32516, G32518, G32520, G32523, G32524, G32537, G32541, G32565, G32566, G32568, G32571, G32578, G32644, G32647, G32649, G32650, G32774, G32776, G32778, G32779, G32786, G32912, G32914, G32915 and G32922; Part Number: 750733, Serial Numbers: G29230, G29891, G30029, G30171, G30172, G30174, G30180, G30322, G30325, G30661, G30666, G30799, G30801, G31007, G31010, G31307, G31310, G315881, G32292, G32535, G32771, G32773, G32918, G-32924, G-32976, G329821, G33085, G33086, G33090, G33361, G33874, G33876, G34118, G34128 and G34232.
510(k) numbers
["K920742","K943581"]
PMA numbers
not on file
Product code
IZO
Firm FEI
1518293
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register