Skip to the content

FDA Medical Device Recalls (openFDA)

BioFire Diagnostics, LLC: BioFire BCID2 Panel, REF RFIT-ASY-0147, For FilmArray Systems, CE, Blood Culture Identification Panel in vitro diagnostic, Rx Only, UDI: (01)00815381020338

Recall number
Z-1903-2021
Recalling firm
BioFire Diagnostics, LLC
Product
BioFire BCID2 Panel, REF RFIT-ASY-0147, For FilmArray Systems, CE, Blood Culture Identification Panel in vitro diagnostic, Rx Only, UDI: (01)00815381020338
Status
Terminated
Initiated
2021-05-10
Posted
not on file
Terminated
2022-12-15
Reason
Elevated rates of false negative results may occur when using blood culture panels due to issues in the manufacturing process. Pouches contained OPP from FilmArray Blood Culture Identification Panel (BCID) instead of OPP from BCID2 Panel.
Root cause
Process control
Action
On 5/10/21, BioFire distributed a "Urgent: Medical Device Correction" letter to all consignees via: E-Mail. In addition to providing consignees with a recall notification, the firm ask consignees to take the following actions: 1 Immediately examine your inventory for product identified in this recall (BCI 2 Panel Pouch Lot # 11YA20/ Kit Lot # 2039020). 2. Discontinue use and discard any remaining product in your possession that is subject to this recall. BioFire will replace the product at no charge in accordance with Biofire's Standard limited Warranty 3.Confirm the amount of pouches scrapped (if any) on the attached acknowledgement of receipt form. 4. If you may have further distributed this product, please identify any recipients and notify them at once of this product recall. 5. Please complete the accompanying Field Action acknowledgement of receipt Form and return to BioFire so that BioFire may acknowledge your receipt of this notification. 6. If you have any questions or concerns, please contact our customer support department at support@biofiredx.com or via phone at 1.800.735.6544. 7. Adverse events may also be reported to FDA's MedWatch Adverse Event Reporting program.
Quantity
870 kits
Distribution
US nationwide distribution
Lot, serial or model codes
Pouch Lot# 11YA20/ Kit Lot# 2039020. UDI: (01)00815381020338
510(k) numbers
["K193519"]
PMA numbers
not on file
Product code
PEN
Firm FEI
3002773840
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register