FDA Medical Device Recalls (openFDA)

Biomet, Inc.: REF 906764, CURVTEK EYE NEEDLE, 12 MM LARGE, STAINLESS STEEL GAMMA STERILIZED, PKG/3, BIOMET SPORTS MEDICINE, 56 EAST BELL DRIVE, P.O. BOX 587, WARSAW, IN 46581 USA, STERILE After the Curvtek device is used to drill a "C" shaped tunnel through the patient's bone, the device is used to pass suture through the tunnel so that soft tissue may be tied down directly onto the bone.

Recall number
Z-1903-2011
Recalling firm
Biomet, Inc.
Product
REF 906764, CURVTEK EYE NEEDLE, 12 MM LARGE, STAINLESS STEEL GAMMA STERILIZED, PKG/3, BIOMET SPORTS MEDICINE, 56 EAST BELL DRIVE, P.O. BOX 587, WARSAW, IN 46581 USA, STERILE After the Curvtek device is used to drill a "C" shaped tunnel through the patient's bone, the device is used to pass suture through the tunnel so that soft tissue may be tied down directly onto the bone.
Status
Terminated
Initiated
2011-03-10
Posted
2011-04-06
Terminated
2011-07-27
Reason
The firm initiated a recall due to the discovery that the devices were not sterilized, as required, prior to distribution.
Root cause
Process control
Action
The firm, BIOMET, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated March 10, 2011 to all customers. The letter described the product, problem and actions to be taken. The letter noted that -Use of a non-sterile Curvtek Eye Needle may lead to patient infection, possibly requiring a revision procedure. The customers were instructed to immediately locate, discontinue use of the product , remove the product from circulation and return the product to Biomet; carefully follow the instructions on the enclosed FAX BACK RESPONSE FORM; complete and return the response form via fax to 574-372-1683 prior to return of product; use priority carrier for shipment, and please confirm receipt of this notice by calling 800-348-9500, extension 3755 or 3756. The firm also informed physicians that had previously utilized the device to monitor patients for infection and provide treatment as appropriate. Questions related to this notice should be directed to 574-371-3755 or 574-371-3756, Monday through Friday, 8am to 5pm.
Quantity
35
Distribution
Worldwide distribution: USA including states of: IN and KY; and countries of: Canada and The Netherlands.
Lot, serial or model codes
Part Number 906764, Lots 748580, 748730 and 932650.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAB
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register