Hill-Rom Manufacturing, Inc.: CLINITRON UP-LIFT" Air Fluidized Therapy Units. Intended to prevent and/or treat pressure ulcer development and wound deterioration in patients who have a significant risk of developing these problems and who generally also have one or more of the following conditions: immobility, poor nutrition, diminished level of consciousness, reduced subcutaneous tissue or multi-system failure.

Recall number
Z-1903-2009
Recalling firm
Hill-Rom Manufacturing, Inc.
Product
CLINITRON UP-LIFT" Air Fluidized Therapy Units. Intended to prevent and/or treat pressure ulcer development and wound deterioration in patients who have a significant risk of developing these problems and who generally also have one or more of the following conditions: immobility, poor nutrition, diminished level of consciousness, reduced subcutaneous tissue or multi-system failure.
Status
Terminated
Initiated
2009-06-24
Posted
2009-08-25
Terminated
2012-07-11
Reason
Incorrectly repaired power cord wires can potentially overheat resulting in smoldering sound foam in the base of the therapy bed.
Root cause
Equipment maintenance
Action
Hill-Rom issued an "Urgent Field Safety Notice" letter dated February 2, 2009 addressed to Facility Risk Managers and Facility Administrators. The notice detailed the issue and instructed users on actions to be taken. For further information, please contact Hill-Rom Customer Service at 1-800-638-2546.
Quantity
625 units (total both units)
Distribution
Worldwide Distribution -- United States, Abu Dhabi, Brazil, Canada, Chile, Columbia, Mexico, Newfoundland, Puerto Rico, Saudi Arabia, Singapore, Taiwan, Turkey and Venezuela.
Lot, serial or model codes
All serial numbers.
510(k) numbers
["K920112"]
PMA numbers
not on file
Product code
INX
Firm FEI
1045510
Firm city
Charleston
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register