FDA Medical Device Recalls (openFDA)
Medtronic Neuromodulation: IsoMed Refill Kit, Model 8553, Lot Number 60538731, is intended for use in refilling Medtronic IsoMed pumps. Contents: Two 22-gauge needles, extension tubing set with y-connector and clamp, Filter, 10-ml syringe, 60-ml lidded syringe, Fenestrated drape, Template; Use by Date May 10, 2009; Medtronic, Inc., Minneapolis, MN. Product is packaged in a sterile container.
- Recall number
- Z-1903-2008
- Recalling firm
- Medtronic Neuromodulation
- Product
- IsoMed Refill Kit, Model 8553, Lot Number 60538731, is intended for use in refilling Medtronic IsoMed pumps. Contents: Two 22-gauge needles, extension tubing set with y-connector and clamp, Filter, 10-ml syringe, 60-ml lidded syringe, Fenestrated drape, Template; Use by Date May 10, 2009; Medtronic, Inc., Minneapolis, MN. Product is packaged in a sterile container.
- Status
- Terminated
- Initiated
- 2008-02-04
- Posted
- 2008-09-17
- Terminated
- 2010-10-13
- Reason
- Medtronic is retrieving all unused packages of ISOMED Refill Kit Model 8553, Lot 60538731. This lot of product was released using incorrect endotoxin specification limits. Endotoxin contamination can potentially lead to fever, malaise, respiratory distress, septic shock, and death.
- Root cause
- Process control
- Action
- Consignees were notified by Medtronic field personnel, either by personal visits or telephone, beginning 2/4/08. This included the 8 additional consignees located outside of the U.S. Medtronic advised customers not to use any unused packages in their inventory and that a field representative would assist them in returning the product for replacement and/or credit. Upon completion of each follow-up, the field representatives completed and returned a Reply Card documenting the notification. If you have questions or comments, contact Medtronic Neuromodulation Technical Services at 1-800-707-0933.
- Quantity
- 1982 kits
- Distribution
- Worldwide -- including the U.S. (AK, AL, AR, AZ, CA, CT, DC,DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and WV) and Argentina, Australia, and Hong Kong.
- Lot, serial or model codes
- Lot 60538731, Use by Date is May 10, 2009.
- 510(k) numbers
- not on file
- PMA numbers
- ["P990034"]
- Product code
- LKK
- Firm FEI
- 2182207
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28