Skip to the content

FDA Medical Device Recalls (openFDA)

Adler MicroMed, Inc.: Sapi Med sterile disposable colorectal anoscopes and proctoscopes. REF/Product Description: A.4083/THE BEAK - Surgical proctoscope. Disposable & sterile. Transparent. Self light; A.4519/MPR Short - Multi Purpose Rectoscope. 190mm long. Sterile & Disposable; A.4086/EPOFLIER - Surgical kit made of a 34,3mm wide anoscope and an opened retractor named GULL. Disposable & sterile; A.4084/THE BEAK Diagnostic- Disposable proctoscope. Transparent. Sterile & Self lighting; A.4522/ MPR - Multi Purpose Rectoscope. 220mm long. Sterile & Disposable

Recall number
Z-1901-2021
Recalling firm
Adler MicroMed, Inc.
Product
Sapi Med sterile disposable colorectal anoscopes and proctoscopes. REF/Product Description: A.4083/THE BEAK - Surgical proctoscope. Disposable & sterile. Transparent. Self light; A.4519/MPR Short - Multi Purpose Rectoscope. 190mm long. Sterile & Disposable; A.4086/EPOFLIER - Surgical kit made of a 34,3mm wide anoscope and an opened retractor named GULL. Disposable & sterile; A.4084/THE BEAK Diagnostic- Disposable proctoscope. Transparent. Sterile & Self lighting; A.4522/ MPR - Multi Purpose Rectoscope. 220mm long. Sterile & Disposable
Status
Terminated
Initiated
2021-04-23
Posted
not on file
Terminated
2022-10-04
Reason
Due to issue with product sterility. Sterilization service provider had falsified records.
Root cause
Process control
Action
On 04/23/2021, the Recalling Firm communicated an URGENT: MEDICAL RECALL message via a telephone calls to customers and followed up email with a notification letter, informing them that the affected products that are labeled as sterile may not be sterile and the products could affect the sterile field and cause patient infections. Customers are instructed to: 1) Segregate and quarantine the affected products and wait for further instructions for disposition to follow, and 2) Complete and return the attached Acknowledgement and Receipt Form and indicate the quantities and location of the recalled products.
Quantity
1,858 devices
Distribution
US distribution: FL, IL, MA, and PA.
Lot, serial or model codes
REF Numbers/Lot Numbers: REF A.4083 /Lot Numbers: 18D004l, 19C010l, 19l018l, 20l027l; REF A.4084/Lot Number: 19l015l; REF A.4519/Lot Numbers: 18l013l, 18l012l, 19D002l, 19F012l; REF A.4522/Lot Numbers: 18G003l, 19K023l; REF A.4086/Lot Number: 18F027l;  Sterilization Batch Numbers: 18042311; 18092611; 18071211_1; 19040512; 18073112; 19092713; 19120612; 19063012; 20102212; 19063012;
510(k) numbers
["K070913","K070915"]
PMA numbers
not on file
Product code
KDM
Firm FEI
3011412640
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register