Skip to the content

FDA Medical Device Recalls (openFDA)

Silex Medical, LLC: Signature Laparoscopic Instruments, 20mm Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5881033.

Recall number
Z-1900-2025
Recalling firm
Silex Medical, LLC
Product
Signature Laparoscopic Instruments, 20mm Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5881033.
Status
Open, Classified
Initiated
2024-01-09
Posted
2025-06-02
Terminated
not on file
Reason
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
Root cause
Process control
Action
Silex formally notified customers on 01/09/2024. Instructions to consignees included to inspect all instruments that they received from the affected lots for the defect. If any instruments had the issue, the customers were instructed to send it back to their facility under an RGA.
Quantity
2 units
Distribution
US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.
Lot, serial or model codes
Model/REF Number: 5881033. UDI-DI: B33158810330. Lot Number: M28188. Serial Numbers: 0107, 0108.
510(k) numbers
["K914910"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3005110411
Firm city
Southington
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register