Skip to the content

FDA Medical Device Recalls (openFDA)

McKesson Provider Technologies - Medical Imaging Group: Horizon Cardiology Hemo Monitoring System A programmable diagnostic computer intended for acquiring digitizing, storing, displaying , and reviewing hemodynamic data for use in hospital cardiac catherization laboratories.

Recall number
Z-1899-2010
Recalling firm
McKesson Provider Technologies - Medical Imaging Group
Product
Horizon Cardiology Hemo Monitoring System A programmable diagnostic computer intended for acquiring digitizing, storing, displaying , and reviewing hemodynamic data for use in hospital cardiac catherization laboratories.
Status
Terminated
Initiated
2010-01-27
Posted
2010-06-24
Terminated
2010-06-25
Reason
McKesson has identified a potential hardware configuration problem in some Horizon Cardiology Hemo systems which may cause delay and/or loss of the patient's physiological parameters on the Horizon Cardiology Hemo screen and on the patient record.
Root cause
Software design
Action
The firm, McKesson Israel Ltd., sent notification letter dated January 27, 2010 to customers. The letter describes the product, problem and action to be taken by customers and the firm. McKesson's will be contacting the customer to coordinate replacement of the sound component of the Horizon Cardiology Hemo system and will provide the additional sound adapter for all affected systems at your facility at not cost. The customers were instructed to pleasel immediately conduct the following instructions to disable sound operations-1) Open McKesson Management Console and locate the Windsurfer>Monitor Settings; 2) Uncheck "Enable QRS beep" in the sounds section; 3) Locate Windsufer>Hardware Settings in McKesson Management Console, and 4) Uncheck "Enable QRS beep" in the Analog Output, QRS Beep Section. If you have any questions, please do not hesitate to contact Ronen Gans directly at (973) 355-9900.
Quantity
22 devices
Distribution
Nationwide distribution: CA, CT, FL, LA, NC, TN, WA, and WV.
Lot, serial or model codes
Serial numbers: 2008052847, 2008072107, 2008092609, 2008092608, 2009010800, 2008121714, 2008101609, 2008101608, 2008101618, 2008110391, 2008110390, 2008110389, 2009020226, 2009020227, 2009030603, 2009030602, 2009030601, 2009110399, 2008110398, 2008051520, 2008051519, 2008051518.
510(k) numbers
["K050561"]
PMA numbers
not on file
Product code
DQK
Firm FEI
3005157178
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register