FDA Medical Device Recalls (openFDA)

FUJIFILM Healthcare Americas Corporation: FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33

Recall number
Z-1898-2024
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33
Status
Open, Classified
Initiated
2024-04-11
Posted
2024-05-24
Terminated
not on file
Reason
Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
Root cause
Component design/selection
Action
FUJIFILM Healthcare Americas Corporation (FUJIFILM) issued Urgent -FIrst Notice letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: 1)Please complete and return the form attached on page 3 of this communication. Providingthe response with the information requested is essential for ensuring appropriate action istaken 2)Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to open a serviceticket. 3)Until your upgrade is scheduled and completed, you can continue to safely use FDRVisionary Suite, however, inspect the bolts prior to use. Please contact our field action coordinator, Chaitrali Kulkarni, via email at hcusregulatoryaffairs@fujifilm.com, if you have any further questions regarding this field action.
Quantity
18 units
Distribution
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
Lot, serial or model codes
UDI-DI: 04540217052226 04540217057450 S/N: MP95A9F6A001, MP95A9F74002, MP95A9F74001, MP95AA983002, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA993001, MP95AA996002, MP95AA996001, MP95AA997001, MP95A9F6B001, MP95A9F76001, MP95AA983001, MP95AA988002, MP95AA98A001, MP95AA993002
510(k) numbers
["K152294"]
PMA numbers
not on file
Product code
KPR
Firm FEI
1000513161
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register