Bio-Rad Laboratories Inc: Device brochure, labeled in part : BIO -RAD***L70242709 VARIANT TM II TURBO***Hemoglobin A 1C***Cartridge Set, Cartridge Set, Analytical Cartridge, Guard Cartridge Insert***UNITED STATES, Bio-Rad Laboratories, Inc. Hercules, CA 94547***" Integrated method for sample preparation, separation and the percent determination of percent hemoglobin A 1 c in EDTA human whole blood.

Recall number
Z-1898-2011
Recalling firm
Bio-Rad Laboratories Inc
Product
Device brochure, labeled in part : BIO -RAD***L70242709 VARIANT TM II TURBO***Hemoglobin A 1C***Cartridge Set, Cartridge Set, Analytical Cartridge, Guard Cartridge Insert***UNITED STATES, Bio-Rad Laboratories, Inc. Hercules, CA 94547***" Integrated method for sample preparation, separation and the percent determination of percent hemoglobin A 1 c in EDTA human whole blood.
Status
Terminated
Initiated
2009-12-08
Posted
2011-04-06
Terminated
2011-06-29
Reason
Wrong lot number was found on the Hemoglobin Cartridge Insert . The insert did not match the lot number to the device cartridge or the device cartridge label or on the Update Kit CD.
Root cause
Other
Action
The firm, Bio-Rad, sent a "MEDICAL DEVICE CORRECTION" letter dated November 29, 2009 to all customers by Federal Express. The letter described the product, problem and action to be taken. The customers were instructed to do the following: 1) If their Lot No. LV90170B VARIANT II TURBO Hemoglobin A(10) Cartridge Set includes an insert referencing Lot No. 80080, please destroy the insert and use the new insert enclosed with the letter. 2) If their Lot No. LV90170B VARIANT II TURBO Hemoglobin A(10) Cartridge Set includes an insert referencing Lot No. 90170, they may continue using the insert. 3) Complete and return the attached CUSTOMER MEDICAL DEVICE CORRECTION RESPONSE FORM via fax to: Bio-Rad CSD Regulatory Affairs Department at (510) 741-3954. If you have any questions, please call 510-724-7000 or contact your regional Bio-Rad office.
Quantity
51 kits (Lot # 70292242; 45 kits( Lot # 70292318)
Distribution
Worldwide distribution: USA including states of: CA, CT, GA, IL, IN, MA, ME, MS, NH, NV, OH, PA, TX, VA and WV; and countries including: France, Hong Kong, Italy, Korea and New Zealand.
Lot, serial or model codes
Model Number(s): 270-2417 (Kit), with 270-2420 (Cartridge set) Code(s): LCP Listing Number(s): DO48707 for CDM 3.6T software Expiration date(s): Lot 70292242 EXP. Mar 31, 2010 and Lot 70292318 EXP. Jan. 31, 2010
510(k) numbers
["K063400"]
PMA numbers
not on file
Product code
LCP
Firm FEI
2915274
Firm city
Hercules
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register