Angiodynamics, INC: SmartPort CT Port Access System, Catalog number CT80STPD, AngioDynamics Incorporated The SmartPort CT Port Access System is indicated for any patient requiring repeated access of the vascular system or other selected body site, for delivery of medications, nutritional supplementation, fluids, blood products, and sampling of blood.

Recall number
Z-1898-2009
Recalling firm
Angiodynamics, INC
Product
SmartPort CT Port Access System, Catalog number CT80STPD, AngioDynamics Incorporated The SmartPort CT Port Access System is indicated for any patient requiring repeated access of the vascular system or other selected body site, for delivery of medications, nutritional supplementation, fluids, blood products, and sampling of blood.
Status
Terminated
Initiated
2009-07-16
Posted
2009-09-01
Terminated
2010-10-27
Reason
The product may contain an incorrect size catheter and introducer.
Root cause
Packaging process control
Action
Notification of the AngioDynamics, Inc. sales force began on July 16, 2009 and they were asked to assist in the return of all affected devices to the Manchester, GA division of AngioDynamics, Inc. Urgent Recall letters were sent to all customers on July 16, 2009, requesting the completion of a FAX BACK form. Those currently in possession of any remaining devices will be provided a Return Goods Authorization (RGA) number and pre-paid shipping boxes for return of the product to AngioDynamics, Inc. All returned devices will be physically and electronically quarantined until final disposition of the devices. All devices will be inspected for damage to the devices and any damaged devices destroyed (scrapped) via the AngioDynamics procedure for Disposition of Nonconformances, procedure. Non-damaged product may be reworked to remove the incorrect catheter and introducer and replaced with the correct catheter and introducer, repackaged and re-sterilized in accordance with the AngioDynamics procedure for Rework. If necessary, follow-up letters, phone calls or visits by the AngioDynamics sales representatives may be used to complete the effective checks. Level A (100%) of consignees to be contacted). Please contact your local representative or Recall Coordinator at 1-706-846-3126, for assistance. You may e-mail your questions to customerservice@angiodynamics.com.
Quantity
150 kits
Distribution
AL, CA, CO, FL, GA, ID, IL, IN, KS, LA, MI, MO, MN, MS, NC, NY, OH, OR, PA, TN and TX.
Lot, serial or model codes
Lot number: 983533
510(k) numbers
["K081472"]
PMA numbers
not on file
Product code
LJT
Firm FEI
3017892510
Firm city
Manchester
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register