Intuitive Surgical Inc: da Vinci X Surgical System, IS4200 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the IS4000 system and its primary function is to support the instrument arms and camera arm. The system has four Universal Surgical Manipulators (USM) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. There are four USM per da Vinci X surgical system.

Recall number
Z-1897-2020
Recalling firm
Intuitive Surgical Inc
Product
da Vinci X Surgical System, IS4200 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the IS4000 system and its primary function is to support the instrument arms and camera arm. The system has four Universal Surgical Manipulators (USM) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. There are four USM per da Vinci X surgical system.
Status
Terminated
Initiated
2019-04-11
Posted
not on file
Terminated
2024-04-08
Reason
Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.
Root cause
Process control
Action
On 04/11/2019, the firm sent an "Urgent Medical Device Correction" notifications via FedEx to its customers informing them that the Recalling Firm has become aware that specific Surgical Systems arms have the potential to encounter friction in specific instrument arm orientations. As a result, sluggish, jerky, or hesitating motion may be experienced at the instrument tip and/or resistance felt in the hand controls. To correct this issue, a Recalling Firm's representative will schedule a site visit to perform evaluations on their systems. Based on the results of the evaluation, any necessary corrections will be prioritized appropriately. In addition: Customers are informed and instructed that If any of instruments arms exhibit sluggish, jerky, or hesitating motion, or if you feel significant resistance from the hand controls with no interference between arms/instruments, contact the da Vinci Surgical Technical Assistance Team (dVSTAT) as soon as possible. General Acknowledgement 1. Forward this letter to your Risk Manager, OR Director, Biomedical Engineering staff, and other members of your medical staff who perform da Vinci procedures. 2. If available, please log into the da Vinci Online Community Field Action resource to read or complete any requested actions related to this issue, at this link: https://www.davincisurgerycommunity.com/ 3. In the event the da Vinci online resource cannot be used, complete the attached Acknowledgement Form and return it via fax to Intuitive as instructed on the form. 4. Please retain a copy of this letter and the acknowledgement form for your files. representatives will be available by phone to answer any questions related to this Medical Device Correction. An Intuitive representative will schedule a site visit with your hospital to perform the inspection and provide any potential correction activity for the instrument arms of your system(s). If you need further information or support concerning this Medical Devic
Quantity
270 USMs on a combination of 100 unique systems (4 USM per system)
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain, Sweden, and the United Kingdom.
Lot, serial or model codes
UDI: 00886874114216 Device Listing: D295931 Serial Numbers: SL0085, SL0315, SL0317, SL0324, SL0338, SL0350, SL0351, SL0353, SL0354, SL0357, SL0369
510(k) numbers
["K171294"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register