Becton Dickinson & Company: BD FlowSmart Set MiniMed Pro-set Rx Only Distributed by Medtronic MiniMed Inc. 8006464633 8185765555 Legal Manufacturer Becton, Dickinson and Company USA BD Medical-Diabetes Care Becton Dickinson France SAS

Recall number
Z-1897-2017
Recalling firm
Becton Dickinson & Company
Product
BD FlowSmart Set MiniMed Pro-set Rx Only Distributed by Medtronic MiniMed Inc. 8006464633 8185765555 Legal Manufacturer Becton, Dickinson and Company USA BD Medical-Diabetes Care Becton Dickinson France SAS
Status
Terminated
Initiated
2016-12-23
Posted
not on file
Terminated
2017-11-09
Reason
Patients are reporting high blood glucose (HBG) levels while using the MiniMed Pro-set and also noticed the catheter of the infusion set was bent/kinked.
Root cause
Device Design
Action
Becton Dickinson (BD) notified their sole customer Medtronic via e-mail on December 26, 2016. BD recommends that Medtronic notify their customers of the situation. For further questions, please call (201) 847-6800.
Quantity
57,750 units
Distribution
US Distribution to the state of : CA
Lot, serial or model codes
6203871 30-Jun-19 6203860 30-Jun-19 6207535 30-Jun-19 6211649 31-Jul-19 6207537 30-Jun-19 6211650 31-Jul-19
510(k) numbers
["K160651"]
PMA numbers
not on file
Product code
FPA
Firm FEI
2243072
Firm city
Franklin Lakes
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register