Skip to the content

FDA Medical Device Recalls (openFDA)

Apheresis Technologies, Inc.: Owner Manual Modifications for PlasmaPro Plasma Pump Model PP-05. Apheresis Technologies, Inc. Palm Harbor, FL. - No labeling or Packaging. Used with Asahi Plasma Separators to perform therapeutic plasma exchange.

Recall number
Z-1897-2009
Recalling firm
Apheresis Technologies, Inc.
Product
Owner Manual Modifications for PlasmaPro Plasma Pump Model PP-05. Apheresis Technologies, Inc. Palm Harbor, FL. - No labeling or Packaging. Used with Asahi Plasma Separators to perform therapeutic plasma exchange.
Status
Terminated
Initiated
2009-04-01
Posted
2009-08-27
Terminated
2010-07-06
Reason
Apheresis Technologies, Inc. is taking the following Corrective Action: Modification of the Owners Manual Only for the PlasmaPro Plasma Pump Model PP-05. Operator's Manual will modify tubeset loading instructions so as to reduce possible loading error which might result in the tube being pulled into the pump head.
Root cause
Labeling design
Action
PlasmaPro Plasma Pump Operators Manual was modified and distributed to firms consignees. Firms received written notification of re-issuance of Operator's Manual.
Quantity
7 Owner's Manuals
Distribution
Nationwide Distribution -- Florida, Colorado and Utah.
Lot, serial or model codes
Model PP-05; all units sold.
510(k) numbers
["K961137"]
PMA numbers
not on file
Product code
FIR
Firm FEI
1000113717
Firm city
Palm Harbor
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register