Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Medical Corporation: SurFlo IV Catheter

Recall number
Z-1896-2017
Recalling firm
Terumo Medical Corporation
Product
SurFlo IV Catheter
Status
Terminated
Initiated
2015-08-05
Posted
not on file
Terminated
2017-05-01
Reason
According to the Terumo (TMC) this device is used to access the human peripheral vascular system for short term use (less than 30 days) for the purpose of infusing fluids into patients. A small percentage of the catheters in the affected lots may have damage to the ETFE catheter which may allow the catheter to break during withdrawal.
Root cause
Process control
Action
Terumo sent an Urgent Product Bulletin letter dated August 3, 2015. The letter identified the product the problem and the action needed to be taken by the customer. 1. CUSTOMERS MUST IMMEDIATELY DISCONTINUE USE, AND QUARANTINE ANY AVAILABLE STOCK, OF THE LISTED PRODUCT CODES. 2. DISTRIBUTORS MUST NOT DISTRIBUTE ANY OF THE LISTED PRODUCT CODES, QUARANTINE ANY AVAILABLE STOCK AND SEND THIS RECALL BULLETIN IMMEDIATELY TO EACH CUSTOMER THAT RECEIVED THESE PRODUCTS TO INFORM THEM OF THIS ISSUE. Terumo is taking this action to assure the safety of our customers and their patients even though it may cause a significant inconvenience. If this creates a critical shortage, please contact us immediately so we may try to provide assistance in procuring replacement or alternative product. For further questions, please call (410) 392-7226
Quantity
2,478,200
Distribution
US Distribution. This product has been sold to both distribution companies with multiple locations and directly to end users/hospitals in US.
Lot, serial or model codes
PRODUCT CODE  3SR-OX2225CA 3SR-OX2419CA 3SR-OX2225CA 3SR-OX2419CA SR-TL2419M 3SR-OX2419CA 3SR-OX2225CA SR-TL2225M 3SR-OX2225CA 3SR-OX2419CA 3SR-OX2225CA 3SR-OX2419CA SR-TL2419M 3SR-OX2225CA 3SR-OX2225CA 3SR-OX2225CA 3SR-OX2225CA SR-TL2225M 3SR-OX2225CA 3SR-OX2225CA 3SR-OX2225CA 3SR-OX2419CA SR-TL2419M 3SR-OX2419CA 3SR-OX2419CA 3SR-OX2419CA  LOT NUMBER  RF2727 RF2727 RG0227 RG0927 RG0927 RG1627 RG2027 RG2027 RG2327 RH0727 RH1427 RH1427 RH1427 RH2127 RH2827 RH3127 RK0427 RK0427 RK1127 RK1827 RK2527 RL0227 RL0227 RL0827 RL1527 RL2227
510(k) numbers
["K891087"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
1118880
Firm city
Elkton
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register