FDA Medical Device Recalls (openFDA)
Respironics, Inc.: Trilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. Provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- Recall number
- Z-1896-2010
- Recalling firm
- Respironics, Inc.
- Product
- Trilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. Provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- Status
- Terminated
- Initiated
- 2010-06-10
- Posted
- 2010-06-24
- Terminated
- 2010-11-18
- Reason
- It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event.
- Root cause
- Nonconforming Material/Component
- Action
- Philips Respironics has contacted all of its consignees by phone and instructed them to remove the potentially affected external battery packs from their inventory and any that may be on their patients/end users. These consignees have been instructed to quarantine batteries they have removed away from flammable materials. Product return/replacement instructions were provided. For further information, contact Philips Respironics at 1-877-387-3311.
- Quantity
- 23 units
- Distribution
- Worldwide Distribution -- United States (MT, IL, IA, NY, NC, TX, CA and FL), France, South Korea and Turkey.
- Lot, serial or model codes
- Serial Numbers: TV110051146, TV110051147, TV110051347, TV110051335, TV110050717, TV110052245, TV110051033, TV110051130, TV110050733, TV110051133, TV110051336, TV110051144, TV110051241, TV110051236, TV110051344, TV110051333, TV110051338, TV110051340, TV110051352, TV110051348
- 510(k) numbers
- ["K083526"]
- PMA numbers
- not on file
- Product code
- CBK
- Firm FEI
- 2518422
- Firm city
- Murrysville
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28