Skip to the content

FDA Medical Device Recalls (openFDA)

Mirion Technologies (Capintec), Inc.: Captus 4000e Thyroid Uptake System with Well and 2 inch shielding. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.

Recall number
Z-1894-2024
Recalling firm
Mirion Technologies (Capintec), Inc.
Product
Captus 4000e Thyroid Uptake System with Well and 2 inch shielding. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.
Status
Open, Classified
Initiated
2024-04-22
Posted
2024-05-24
Terminated
not on file
Reason
Complaints of unexpected detachment of the collimator have been reported.
Root cause
Device Design
Action
Consignees were sent an URGENT SAFETY NOTICE via email, dated 4/1/24. Consignees are asked to review the provided notice, share the notice with ass affected personnel, instruct personnel to comply with the provided instructions for safe operation and how to perform a retention plate inspection, and to contact Capintec Technical Support if the inspection shows that the collimator is not properly attached. Capintec Technical Support can be reached by phone at 1-800-631-3826 or at 1-201-825-9500, or by email at capintecsupport@mirion.com. Consignees are also asked to confirm receipt of the notification utilizing the provided QR Code or form.
Quantity
1,275 units
Distribution
Worldwide - US Nationwide distribution and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, FRANCE, GERMANY, GUAM, HONDURAS, HUNGARY, INDIA, ISRAEL, ISREAL, ITALY, JAPAN, KOREA, LATVIA, MALAYSIA, NETHERLANDS, NEW ZEALAND, PAKISTAN, PHILIPPINES, PHILIPPINES , PHILLIPINES, PUERTO RICO, QATAR, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TAIWAN , THAILAND, TURKEY, UNITED KINGDOM, VIETNAM.
Lot, serial or model codes
Lot Code: Model No: 5430-30154; UDI 0859942006423; S/N range 940001 through 941398.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IZD
Firm FEI
2518443
Firm city
Florham Park
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register