FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry System Computerless, Catalog #700-7177-001. The iChemVELOCITY automated urine chemistry system is an in-vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in a iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems.
- Recall number
- Z-1894-2018
- Recalling firm
- Beckman Coulter Inc.
- Product
- iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry System Computerless, Catalog #700-7177-001. The iChemVELOCITY automated urine chemistry system is an in-vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in a iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems.
- Status
- Terminated
- Initiated
- 2018-03-29
- Posted
- 2018-04-27
- Terminated
- 2019-06-27
- Reason
- The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.
- Root cause
- Process control
- Action
- The recalling firm, Beckman Coulter/IRIS, sent an "URGENT MEDICAL DEVICE RECALL" letter dated dated 3/26/2018 via email on 3/29/2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: -Follow the on-board strip stability claims per Table 1 to determine the frequency at which the strips need to be changed. -Whenever the Strip Provider Module is accessed (i.e., opened or when loading strips), remove and discard the remaining on-board strips and desiccant before replenishing. -It is suggested that for the tests performed since the installation of the iChemVELOCITY/SPM housing assembly, consult with your Laboratory Director to determine whether a retrospective review of results is clinically warranted. -Complete and return the CUSTOMER RESPONSE FORM within 10 days by email to: Regulatory.Notifications@beckman.com, Please include FA-33476 iChemVELOCITY Automated Urine Chemistry System in e-mail subject field or by mail to:Beckman Coulter, Inc., Attn: Regulatory Affairs Mail Code 31-B06 11800 S.W. 147th Avenue, P O BOX 169015, Miami, FL 33116-9015. -Beckman Coulter will contact you by phone to schedule a site visit for replacing the SPM on your instrument. -Beckman Coulter will be implementing process improvements to prevent this issue from recurring. If you have any questions regarding this notice, please contact: - From our website: http://www.beckmancoulter.com - By phone: call 800-854-3633 in the United States and Canada. -Outside the United States and Canada, contact your local Beckman Coulter representative.
- Quantity
- 6 devices
- Distribution
- US Distribution to states of: AL, IA, GA, and FL.
- Lot, serial or model codes
- Catalog 700-7176-001 - Serial numbers V04494 and CV04015; Catalog 700-7177-001 - Serial numbers CV04012, CV04013, CV04014, and CV04017.
- 510(k) numbers
- ["K101852"]
- PMA numbers
- not on file
- Product code
- KQO
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28