FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem VELOCITY automated urine chemistry system is an in vitro diagnostic deice used to automate the urine chemistry analysis profile using iChem VELOCITY Urine Chemistry Strips. The iChem VELOCITY can be used as a stand alone-system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterase, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChem VELOCITY strips are intended for use with the iCHEM VELOCITY analyzer. In particular they are not intended for visual reading. The iChem VELOCITY test strips are not intended for visual reading. The iChem VELOCITY is not intended to be used as a Point of Care (POC) analyzer. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function. Tests performed using the iCHEM VELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.
- Recall number
- Z-1893-2018
- Recalling firm
- Beckman Coulter Inc.
- Product
- iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem VELOCITY automated urine chemistry system is an in vitro diagnostic deice used to automate the urine chemistry analysis profile using iChem VELOCITY Urine Chemistry Strips. The iChem VELOCITY can be used as a stand alone-system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterase, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChem VELOCITY strips are intended for use with the iCHEM VELOCITY analyzer. In particular they are not intended for visual reading. The iChem VELOCITY test strips are not intended for visual reading. The iChem VELOCITY is not intended to be used as a Point of Care (POC) analyzer. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function. Tests performed using the iCHEM VELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.
- Status
- Terminated
- Initiated
- 2017-12-04
- Posted
- not on file
- Terminated
- 2023-05-30
- Reason
- A subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false negative erroneous results for leukocytes.
- Root cause
- Mixed-up of materials/components
- Action
- The firm initiated the recall by letter on 12/04/2017. The letter identified the affected product, problem and the actions to be taken. The firm requested destruction of the recalled product.
- Quantity
- 14340 units
- Distribution
- Worldwide Distribution and US Nationwide Distribution
- Lot, serial or model codes
- Lot #7212154 A
- 510(k) numbers
- ["K101852"]
- PMA numbers
- not on file
- Product code
- LJX
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28