Skip to the content

FDA Medical Device Recalls (openFDA)

Ziehm Imaging Inc: Ziehm Imaging solo FD Mobile Fluoroscopic C-arm

Recall number
Z-1893-2017
Recalling firm
Ziehm Imaging Inc
Product
Ziehm Imaging solo FD Mobile Fluoroscopic C-arm
Status
Terminated
Initiated
2017-02-23
Posted
not on file
Terminated
2019-12-11
Reason
During Ziehm Imaging GmbH FDA inspection, the label for the Collimator and X-ray Control panel was missing the manufacturer date as part of the identification for certified component
Root cause
Labeling design
Action
Ziehm Imaging will bring defect into compliance: Customer Precautions: Therefore, as there is no identifiable risk to patients, users or third parties, the manufacturer Ziehm Imaging GmbH advises customers they may continue to operate the affected device until the labeling materials can be remediated. Please alert all staff members working with the device to this notification. Remedy: 1. You will contact customers and initiate the actions to correct the failure to comply issues. 2. You will affix the identification labels to the collimator and x-ray control panel. 3. You will replace the existing user manual with one that reflects the reference images of the labels with manufacturer date. If there are any questions or concerns regarding this product safety issue, please do not hesitate to contact our National Service Team for further information at Ziehm Imaging, Inc. Toll-Free: (866) 949-4346, Fax: (866) 839-7893 or E- Mail: customerservice@ziehm.com
Quantity
US - 2 systems
Distribution
US Distribution to the states of : FL and CA. No devices listed in this recall where distributed any foreign /military/government consignees from the U.S.A.
Lot, serial or model codes
S/N 52199, 52108
510(k) numbers
["K161976"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2027299
Firm city
Orlando
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register