Skip to the content

FDA Medical Device Recalls (openFDA)

Phillips Nuclear Medicine: BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc., Cleveland, OH, 3/8" tilt, model number 4535604 34811, 3/4"tilt, model number 4535674 34801, and 3/8" fixed, model number 4535604 34781, all catalog number 882480 Medical device-- intended to produce images depicting anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.

Recall number
Z-1892-2009
Recalling firm
Phillips Nuclear Medicine
Product
BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc., Cleveland, OH, 3/8" tilt, model number 4535604 34811, 3/4"tilt, model number 4535674 34801, and 3/8" fixed, model number 4535604 34781, all catalog number 882480 Medical device-- intended to produce images depicting anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.
Status
Terminated
Initiated
2009-07-02
Posted
2009-09-16
Terminated
2010-04-12
Reason
Unexpected movement-- if the drive assembly is severely compromised, the detector may slide to its hardware limit.
Root cause
Process control
Action
Firm initiated its recall notification on July 6, 2009 issuing a field safety notice to its consignees via FedEx and internationally through its personnel. Responses will be tracked for effectiveness. Questions may be directed to Philips Customer Care Center at 1.800.722.9377.
Quantity
288 units
Distribution
Worldwide
Lot, serial or model codes
All units: 3/8" tilt, model number 4535604 34811, 3/4"tilt, model number 4535674 34801, and 3/8" fixed, model number 4535604 34781, all catalog number 882480
510(k) numbers
["K062298"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
Milpitas
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register