FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes ARTHROSCOPY PACK DYNJ45173D

Recall number
Z-1888-2026
Recalling firm
Medline Industries, LP
Product
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes ARTHROSCOPY PACK DYNJ45173D
Status
Open, Classified
Initiated
2026-02-27
Posted
2026-04-17
Terminated
not on file
Reason
Unapproved design changes to the products outside of the 510(k) clearance.
Root cause
Component change control
Action
On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Quantity
270,311 total
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Lot, serial or model codes
DYNJ45173D UDI-DI 10889942485222 (EA) 40889942485223 (CS) LOT 24BBB811
510(k) numbers
not on file
PMA numbers
not on file
Product code
OII
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register