FDA Medical Device Recalls (openFDA)
AtriCure, Inc.: EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The EPi-Sense Guided Coagulation System with VisiTrax, consists of a sterile, single-use, disposable coagulation electrode device (1cm, 2cm, and 3cm sizes provided) intended to be used to coagulate cardiac tissue. The flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency (RF) energy from an Electrosurgical Generator (non-sterile, re-useable) connected through an Instrument Cable (sterile). A temporary sensing electrode feature may be used with an off-the-shelf electrogram recording system with the use of an additional instrument cable (non-sterile). An accessory Cannula may be used to facilitate coagulation device access and visibility of the heart. The EPi-Sense Guided Coagulation System with VisiTrax technology is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, endoscopic, and laparoscopic surgical techniques and may be used for temporary cardiac signal sensing and recording during surgery when connected to a temporary external recording device.
- Recall number
- Z-1887-2019
- Recalling firm
- AtriCure, Inc.
- Product
- EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The EPi-Sense Guided Coagulation System with VisiTrax, consists of a sterile, single-use, disposable coagulation electrode device (1cm, 2cm, and 3cm sizes provided) intended to be used to coagulate cardiac tissue. The flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency (RF) energy from an Electrosurgical Generator (non-sterile, re-useable) connected through an Instrument Cable (sterile). A temporary sensing electrode feature may be used with an off-the-shelf electrogram recording system with the use of an additional instrument cable (non-sterile). An accessory Cannula may be used to facilitate coagulation device access and visibility of the heart. The EPi-Sense Guided Coagulation System with VisiTrax technology is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, endoscopic, and laparoscopic surgical techniques and may be used for temporary cardiac signal sensing and recording during surgery when connected to a temporary external recording device.
- Status
- Terminated
- Initiated
- 2019-05-29
- Posted
- not on file
- Terminated
- 2020-05-04
- Reason
- The sterile package seal for the Epi-Sense Guided Coagulation System may be compromised, and has a worst-case reasonable harm of infection if used.
- Root cause
- Process control
- Action
- On May 29, 2019, the firm notified customers of the recall via Urgent Advisory Notice. Customers were asked to take the following actions: " Immediately quarantine any affected product as identified within this Advisory Notice. " Contact AtriCure product complaints by phone at 1-866-349-2342 (select option 6) or e-mail to pcomplaints@atricure.com to request a Return Goods Authorization (RGA). " Return the attached Acknowledgement Form. See Attachment B. The acknowledgement form must be completed and returned even if product is not on hand by selecting the second option. " Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you have any questions, please contact Rob Cantu, Vice President of Quality at (1-513-644-4245) from 9-6pm ET on Mondays - Fridays. You may also contact customer service at (1-866-349-2342) any time of day, your message will be forwarded to Quality Assurance for review promptly. This advisory issue will also be posted on AtriCures website at www.atricure.com/products.
- Quantity
- 497
- Distribution
- Worldwide distribution - US Nationwide distribution, and countries of Germany and Netherlands.
- Lot, serial or model codes
- Lots 70638, 71332, 89208, 89938, 90624
- 510(k) numbers
- ["K142084"]
- PMA numbers
- not on file
- Product code
- OCL
- Firm FEI
- 3011706110
- Firm city
- Mason
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28