Heartware, Inc.: HeartWare Ventricular Assist System: HeartWare Battery, Lithium Ion, 14.8 V, Rx only. Catalog 1650 (US) and 1650, 1650-DE and A00035 (OUS) Product Usage: Used as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare Ventricular Assist System is designed for in-hospital and out-of-hospital settings, including transportation via fixed-wing aircraft or helicopter.
- Recall number
- Z-1887-2016
- Recalling firm
- Heartware, Inc.
- Product
- HeartWare Ventricular Assist System: HeartWare Battery, Lithium Ion, 14.8 V, Rx only. Catalog 1650 (US) and 1650, 1650-DE and A00035 (OUS) Product Usage: Used as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare Ventricular Assist System is designed for in-hospital and out-of-hospital settings, including transportation via fixed-wing aircraft or helicopter.
- Status
- Terminated
- Initiated
- 2016-01-07
- Posted
- 2016-07-08
- Terminated
- 2020-03-27
- Reason
- Heartware Ventricular Assist System Battery Cell experiencing premature depletion.
- Root cause
- Software design
- Action
- The firm, HeartWare, sent an "URGENT MEDICAL DEVICE RECALL" letter dated January 7, 2016, to all their consignees informing them that of the HeartWare HVAD Battery recall (device exchange). The letter described the product, problem and actions to be taken. The customers were instructed to: 1. Promptly complete the attached Acknowledgment Form and return it to HeartWare no later than 30 days from the date of this letter via email to HeartWare7786@stericycle.com or provide verbal Acknowledgement by calling Customer Service line , 1(844)491-7871; or fax to +1 (844)721-3044.. 2. Immediately identify and quarantine affected batteries in hospital inventory, and apply a RECALLED PRODUCT DO NOT USE sticker (provided with this notice) as illustrated below. Replace quarantined inventory AC Adapters and batteries by requesting a Return Goods Authorization (RGA) through a qualified representative or directly through HeartWare Customer Service (via telephone 1-877-367-4823 or email cs@heartware.com). A qualified representative will update the Controller software for affected Controller units in inventory. 3. Identify, collect and quarantine affected product in patient possession. Arrange for current patients to bring their Batteries to a clinic appointment as soon as possible, (at least within the next three months), with urgency dependent upon individual patient circumstances. During their regularly scheduled appointment, a qualified representative will assist with quarantining and replacing affected product. Apply a RECALLED PRODUCT DO NOT USE sticker as illustrated on affected batteries. Replace affected patient product using hospital inventory, and replace consumed hospital inventory by requesting an RGA as described above. 4. Return affected product to HeartWare. Your HeartWare representative will assist you with returning affected product to HeartWare. 5. Complete and return the attached Completion Form no later than 6 months from the date of this let
- Quantity
- 18, 631
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and Washington D.C.
- Lot, serial or model codes
- Serial numbers ranging BAT000001 to BAT199999.
- 510(k) numbers
- not on file
- PMA numbers
- ["P100047"]
- Product code
- DSQ
- Firm FEI
- 3007042319
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28