FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A
- Recall number
- Z-1886-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A
- Status
- Open, Classified
- Initiated
- 2026-02-27
- Posted
- 2026-04-17
- Terminated
- not on file
- Reason
- Unapproved design changes to the products outside of the 510(k) clearance.
- Root cause
- Component change control
- Action
- On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
- Quantity
- 270,311 total
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
- Lot, serial or model codes
- DYNJ35056A UDI-DI 10889942227563 (EA) 40889942227564 (CS) LOT 24ABN639 DYNJ69318A UDI-DI 10195327370756 (EA) 40195327370757 (CS) LOTS 23DMD456 23EMB847 23HMB969 23IMC364 23JME011 23JMI841 23LMC384 24BMD333 24CMB768 24CMJ586 24EMJ527 24GMD970 24JMB408 24JMI594 24KMC774 ***Updated 5/20/26 - The following lots were included in the customer letter but not the FDA submission.*** DYNJ69318A lots 23FMA845 23FMA924 25BMF603
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJK
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28