Stryker Leibinger GmbH & Co. KG: 8000-021-002, Scopis ENT Software with TGS with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.
- Recall number
- Z-1885-2024
- Recalling firm
- Stryker Leibinger GmbH & Co. KG
- Product
- 8000-021-002, Scopis ENT Software with TGS with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.
- Status
- Open, Classified
- Initiated
- 2024-04-04
- Posted
- 2024-05-23
- Terminated
- not on file
- Reason
- Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
- Root cause
- Software Manufacturing/Software Deployment
- Action
- On April 22, 2024 an URGENT MEDICAL DEVICE FIELD ACTION letters were sent to customers. Next Steps: 1. Locate any ENU. Quarantine and discontinue use until software on device has been confirmed. - Locate software version and serial number (see pg. 2 for instructions). - Your Stryker Sales Representatives may help, if needed. 2. Return the enclosed Business Reply Form (BRF) via email to instruments.recalls@stryker.com. - If Scopis ENT Software with TGS version NOVA 3.6.0 RC16 is installed, Stryker will arrange shipment of compliant software. 3. Once compliant software is received and installed, return the enclosed Software Update Form via email to instruments.recalls@stryker.com to confirm completion of the recall action steps. 4. Maintain this notification internally until all required actions have been completed. Please contact Lacey Marshall at 269-270-5950 or Instruments.recalls@stryker.com with questions or concerns. Report any serious adverse events or product quality problems to Stryker Instruments: 1-800-253-3210. Health care professionals and consumers may report serious adverse events or product quality problems with the use of this product to the FDA's MedWatch Adverse Event Reporting program either online, or by fax or phone. Online: ww.fda.gov/Safety/MedWatch/HowToReport/default.htm Fax: (800) FDA-0178 Phone: (800) FDA-1088.
- Quantity
- 705 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Sofware Version: NOVA 3.6.0 RC16, Product Code: 8000-021-002; UDI/DI: 07613327514179; Serial Numbers: 10984, 10110, 10164, 10454, 10762, 10387, 10120, 10252, 10414, 10612, 10408, 10673, 11013, 11032, 10566, 10974, 0310-B, 10326, 10442, 10729, 10681, 10552, 10088, 10476, 10307, 10121, 10371, 0252-B, 10262, 10602, 10613, 10381, 10234, 200774, 10135, 0304-B, 10303, 10257, 10639, 10445, 10684, 10740, 10986, 10413, 10079, 10528, 10565, 10482, 10209, 10574, 10606, 10229, 10214, 10797, 0335-B, 10284, 10465, 10295, 10308, 0282-B, 83B750AD, 0287-B, 10763, 10434, 11025, 10578, 10473, 10596, 0291-B, 10624, 10591, 11092, 10758, 10097, 10294, 0251-B, 10597, 10769, 0309-B, 10765, 0313-B, 0281-B, 10513, 11088, 10057, 10337, 10464, 10621, 10699, 10509, 10421, 10549, 10003, 10499, 10661, 10668, 0292-B, 10416, 0263-B, 10654, 10571, 0329-B, 10934, 10798, 11022, 10610, 10766, 10471, 10437, 10648, 10410, 10403, 10159, 10192, 10372, 10189, 10193, 10642, 10390, 10771, 10202, 10806, 10237, 10296, 0244-B, 0307-B, 10034, 10440, 10704, 10101, 10153, 10098, 10388, 10720, 10943, 10944, 10994, 10695, 10608, 10615, 10054, 10076, 10015, 10089, 10208, 10776, 10780, 11111, 10586, 10759, 10431, 10772, 10376, 10563, 10978, 10989, 10369, 10641, 10796, 10558, 10660, 10071, 10987, 10953, 10982, 10744, 10026, 0325-B, 10959, 11000, 10453, 0332-B, 10055, 10335, 10742, 10649, 10983, 10512, 10958, 10268, 10640, 0280-B, 11075, 11117, 10356, 10429, 10432, 10659, 10677, 10319, 10311, 10332, 10757, 10809, 10523, 10382, 10451, 10495, 10997, 11116, 11018, 10083, 10672, 10775, 10053, 10195, 10330, 10161, 10007, 10328, 10949, 11031, 10781, 10990, 10350, 10750, 10191, 10033, 10056, 10446, 10078, 10516, 10373, 10722, 10136, 10210, 10091, 10255, 10253, 10060, 10427, 10162, 10297, 10731, 10935, 0285-B, 10336, 10760, 10386, 10347, 10286, 10165, 10041, 10456, 10312, 10417, 10500, 10133, 10224, 10031, 10040, 10100, 10784, 10248, 10423, 10215, 10951, 10092, 0260-B, 10185, 10127, 10112, 10790, 10299, 10316, 10435, 10459, 10075, 10479, 10357, 10627, 10732, 10734, 0343-B, 0320-B, 10517, 10455, 10590, 11030, 10383, 10043, 10096, 10789, 10631, 10489, 10791, 10220, 10658, 10348, 10143, 10114, 10625, 10799, 11108, 10103, 10522, 10126, 10066, 10980, 11102, 0318-B, 11005, 10177, 10662, 10166, 10197, 10300, 10406, 10181, 0326-B, 10443, 10331, 10364, 10368, 10592, 10062, 10366, 10349, 10353, 11097, 10155, 10218, 10227, 10230, 10315, 10618, 10021, 10787, 11038, 0271-B, 10365, 10637, 10141, 10525, 10341, 10433, 10696, 0333-B, 10401, 10633, 10211, 10231, 10052, 10794, 10685, 10157, 10213, 10081, 10123, 10090, 10199, 10678, 11020, 10216, 10377, 10424, 10212, 10361, 10059, 10109, 10736, 10344, 10693, 10977, 11004, 10016, 10168, 10171, 10691, 10006, 10458, 10324, 10156, 10190, 10221, 10363, 10475, 0322-B, 0339-B, 10692, 10679, 10082, 10573, 10603, 10267, 10541, 10999, 10169, 10628, 10576, 10345, 10738, 10354, 10238, 0259-B, 10077, 0306-B, 10461, 10172, 10243, 10233, 10327, 0327-B, 11023, 11114, 10450, 10398, 10404, 0303-B, 10400, 10412, 10593, 10072, 10514, 10239, 10069, 10244, 10249, 10761, 10359, 10739, 10352, 10163, 10374, 10698, 10342, 10607, 10635, 10058, 10425, 10419, 10950, 10405, 10422, 10223, 0308-B, 10010, 10594, 10205, 10370, 10474, 10247, 10020, 10174, 10184, 10188, 10128, 11003, 10689, 10701, 10795, 10074, 10266, 10288, 10008, 10203, 10966, 10438, 10955, 10073, 10080, 10333, 10777, 10508, 10325, 10086, 10104, 10469, 10755, 10810, 10671, 10152, 10393, 10657, 10105, 10298, 10524, 10805, 10375, 10402, 10094, 10358, 10498, 10501, 10384, 10391, 10993, 10064, 10175, 10397, 10560, 10570, 10630, 10132, 10579, 10439, 10415, 10306, 10343, 10477, 10323, 10737, 10589, 10304, 10338, 10460, 10598, 10779, 10988, 10346, 11082, 10436, 10362, 10334, 10682, 10653, 10583, 10724, 10725, 10407, 10380, 10409, 10595, 10137, 10656, 10389, 10611, 10638, 11036, 11037, 10577, 10609, 10587, 10626, 10965, 11017, 11112, 10547, 10954, 11029, 11026, 10979, 10985, 11080, 10981, 11099, 10207, 10049, 10580, 10144, 10723.
- 510(k) numbers
- ["K221098"]
- PMA numbers
- not on file
- Product code
- PGW
- Firm FEI
- 3002821504
- Firm city
- Freiburg Im Breisgau
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28