Skip to the content

FDA Medical Device Recalls (openFDA)

Amendia, Inc: 01.4mm x 457mm (18in) K-Wire, Threaded Blunt, Ref 9080B-18T, Qty: 1 per pack, Material: SS 316L, Full Temper, per ASTM F138, Amendia, 1755 West Oak Parkway, Marietta, GA 30062 Guide Wires are used to establish a pathway into the bony pedicle to place implants during spinal surgery

Recall number
Z-1885-2017
Recalling firm
Amendia, Inc
Product
01.4mm x 457mm (18in) K-Wire, Threaded Blunt, Ref 9080B-18T, Qty: 1 per pack, Material: SS 316L, Full Temper, per ASTM F138, Amendia, 1755 West Oak Parkway, Marietta, GA 30062 Guide Wires are used to establish a pathway into the bony pedicle to place implants during spinal surgery
Status
Terminated
Initiated
2017-04-06
Posted
2017-04-25
Terminated
2017-05-30
Reason
Mislabeled guide wires - It was discovered that some Guide Wires labeled as stainless steel were actually manufactured from nitinol.
Root cause
Process design
Action
Amendia sent an Urgent Medical Device Recall Letter dated April 6, 2017, and April 25, 2017, to all affected customers. The letter requested that they check inventory, quarantine product for return to Amendia, and requested a sub-recall. The letter also included a reply form to be returned to Amendia. Customers with questions were instructed to call Amendia Customer Service at 770-575-5224. For questions regarding this recall call 404-610-7215.
Quantity
15 units
Distribution
US distribution to GA, KS & TX
Lot, serial or model codes
Lot Number 135561
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3007970248
Firm city
Marietta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register