Siemens Medical Solutions USA, Inc.: TEE Transducer V5M, when used with the Acuson S2000 Ultrasound System, software versions: VA15, VA15A or VA15B; Model Numbers: 10041461 (Acuson S2000) and 8263703 (V5M Transducer); Product is manufactured and distributed by Siemens Medical Solutions USA, Inc., Mountain View, CA Product is a medical device, a multi-purpose mobile, software controlled diagnostic ultrasound system with and on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. tis function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-mode, MMode, Pulsed(PW) Doppler mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or harmonic imaging and 3d/4d imaging on a flat panel display. The device system is intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The device is a multi frequency, multi plane, phased sector array transducer. In the TEE (transesophageal echocardiography process where the device fails, the transducer is introduced into the esophagus of the patient to obtain images of the heart structure.

Recall number
Z-1884-2011
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
TEE Transducer V5M, when used with the Acuson S2000 Ultrasound System, software versions: VA15, VA15A or VA15B; Model Numbers: 10041461 (Acuson S2000) and 8263703 (V5M Transducer); Product is manufactured and distributed by Siemens Medical Solutions USA, Inc., Mountain View, CA Product is a medical device, a multi-purpose mobile, software controlled diagnostic ultrasound system with and on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. tis function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-mode, MMode, Pulsed(PW) Doppler mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or harmonic imaging and 3d/4d imaging on a flat panel display. The device system is intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The device is a multi frequency, multi plane, phased sector array transducer. In the TEE (transesophageal echocardiography process where the device fails, the transducer is introduced into the esophagus of the patient to obtain images of the heart structure.
Status
Terminated
Initiated
2009-04-07
Posted
2011-04-04
Terminated
2011-04-05
Reason
Device function failure--When performing a TEE exam with the transducer, CW (Continuous Wave) mode is not functional. Due to work flow, this may not be apparent until the transducer is inserted into the patient's esophagus.
Root cause
Software design
Action
Siemens sent a Customer Safety Advisory Notification letter on April 7, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to pass the notice on to all those within their organization who need to be aware of the issue until the corrective action is completed. For questions regarding this recall call 650-969-9112.
Quantity
61 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia, Canada, Switzerland, China, Czech Republic, Germany, Denmark, Great Britain, Hong Kong, Indonesia, India, Italy, South Korea, Norway, Turkmenistan, and Taiwan
Lot, serial or model codes
Serial Numbers: 200190, 200191, 200281, 200287, 200366, 200481, 200519, 200577, 200699, 200738, 200759, 200761, 200830, 200836, 200851, 200855, 200857, 200874, 200878, 200925, 201002, 200909, 200966, 200220, 200187, 200218, 200663, 200524, 200395, 200291, 200850, 200854, 200875, 200965, 200178, 200103, 200104, 200118, 200143, 200183, 200200, 200210, 200686, 200694, 200852, 200889, 200908, 200927, 201005, 200849, 200834, 200835, 200839, 200211, 200696, 200859, 200926, 200660, 200661, 200687, 200719
510(k) numbers
["K082142"]
PMA numbers
not on file
Product code
IYO
Firm FEI
2936884
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register