Skip to the content

FDA Medical Device Recalls (openFDA)

Greatbatch Medical: Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.

Recall number
Z-1882-2016
Recalling firm
Greatbatch Medical
Product
Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.
Status
Terminated
Initiated
2016-04-26
Posted
not on file
Terminated
2018-06-07
Reason
Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.
Root cause
Device Design
Action
Consignees were sent on 4/26/2016 a Greatbatch "Urgent Medical Device Recall" letter dated April 26, 2016. The letter described the product involved in the recall, the reason, and the risk to health. The letter advised consignees to identity and quarantine the product and to complete and return the Filed Recall Response Form. For questions contact your local Greatbatch Medical representative, at 1-763-951-8235, or e-mail FieldActionCenter@Greatbatch.com.
Quantity
2
Distribution
US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
Lot, serial or model codes
6307760001
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWE
Firm FEI
2183787
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register