Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes BARIATRIC DYNJ905153M, DYNJ905153N GENERAL LAPAROSCOPY PACK DYNJ81602 LAPAROSCOPY DYNJ905157K, DYNJ905157L LAPAROTOMY PACK DYNJ46675G MAJOR LITHOTOMY PACK DYNJ81607 MAJOR PLUS PACK DYNJ905160O MINOR GENERAL PACK DYNJ81599

Recall number
Z-1881-2026
Recalling firm
Medline Industries, LP
Product
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes BARIATRIC DYNJ905153M, DYNJ905153N GENERAL LAPAROSCOPY PACK DYNJ81602 LAPAROSCOPY DYNJ905157K, DYNJ905157L LAPAROTOMY PACK DYNJ46675G MAJOR LITHOTOMY PACK DYNJ81607 MAJOR PLUS PACK DYNJ905160O MINOR GENERAL PACK DYNJ81599
Status
Open, Classified
Initiated
2026-02-27
Posted
2026-04-17
Terminated
not on file
Reason
Unapproved design changes to the products outside of the 510(k) clearance.
Root cause
Component change control
Action
On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Quantity
270,311 total
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Lot, serial or model codes
DYNJ905153M UDI-DI 10195327556761 (EA) 40195327556762 (CS) LOT 24CMG485 DYNJ905153N UDI-DI 10198459017797 (EA) 40198459017798 (CS) LOT 25DMC254 DYNJ81602 UDI-DI 10195327123857 (EA) 40195327123858 (CS) LOT 24ABP080 DYNJ905157K UDI-DI 10195327556921 (EA) 40195327556922 (CS) LOTS 24DMF349 24EMC297 DYNJ905157L UDI-DI 10198459017711 (EA) 40198459017712 (CS) LOTS 24FMD083 24HMD400 24IME332 24JMA234 24KMG588 25AMA393 25AMJ541 25BMI202 25CMD909 25CMH371 DYNJ46675G UDI-DI 10195327419066 (EA) 40195327419067 (CS) LOT 24ABS814 DYNJ81607 UDI-DI 10195327123918 (EA) 40195327123919 (CS) LOT 24ABP097 DYNJ905160O UDI-DI 10198459017742 (EA) 40198459017743 (CS) LOTS 24HMG093 24IME714 24JMI924 24LMB556 25AME613 25BMJ076 DYNJ81599 UDI-DI 10195327124021 (EA) 40195327124022 (CS) LOT 24BBB847 ***Updated 5/20/26 - The following lots were included in the customer letter but not in the FDA submission.*** DYNJ905153M lots 24BMH968 24EMG682; DYNJ905153N lots 24HMG087 24IMA442 24IMG245; DYNJ905157K lots 23LME674 24AMH261 24AMJ205 24BME886 24CMD057
510(k) numbers
not on file
PMA numbers
not on file
Product code
FDE
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register