Ventlab LLC: STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide gas analyzer Intended for use in various clinical situations to provide respiratory support in the presence of reversible apnea commonly associated with respiratory arrest. The device should only be used by personnel trained in CPR procedures. The ACE/STAT-Check II is intended for use as an adjunct assessment tool for verification of successful ventilation / intubation in patients > 2.2 kilograms (5lbs.).

Recall number
Z-1880-2016
Recalling firm
Ventlab LLC
Product
STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide gas analyzer Intended for use in various clinical situations to provide respiratory support in the presence of reversible apnea commonly associated with respiratory arrest. The device should only be used by personnel trained in CPR procedures. The ACE/STAT-Check II is intended for use as an adjunct assessment tool for verification of successful ventilation / intubation in patients > 2.2 kilograms (5lbs.).
Status
Terminated
Initiated
2016-03-15
Posted
2016-06-01
Terminated
2016-11-01
Reason
Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.
Root cause
Process control
Action
Ventlab sent an URGENT: MEDICAL DEVICE PRODUCT REMOVAL notification dated March 14, 2016, on March 15, 2016, electronically to all affected customers. Customers were instructed to return the affected product and notify affected consignees of the recall. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this action, please call Ventlab at 1-844-635-5326, or email at PFA@ventlab.com.
Quantity
2430
Distribution
Nationwide distribution to AZ, MS, NY, MO, FL, WI, TX, FL, and CA.
Lot, serial or model codes
Model SC9001B Adult STAT-Check Resuscitation Bag w/ Inflatable Bag Reservoir & Mask Lot numbers 303320 and 303612  Model SC9001C Adult STAT-Check Resuscitation Bag w/ Corrugate Tube Reservoir & Mask Lot number 303400  Model SC9001C-C Adult STAT-Check Resuscitation Bag w/ Corrugate Tube Reservoir, Mask & Straight Connector Lot numbers 303523 and 303791  Model SC8001B Pediatric STAT-Check Resuscitation Bag w/ Inflatable Bag Reservoir & Mask Lot number 303833
510(k) numbers
["K093525","K982215"]
PMA numbers
not on file
Product code
CCK
Firm FEI
1314417
Firm city
Grand Rapids
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register