FDA Medical Device Recalls (openFDA)

Arrow International Inc: Pressure Injectable Arrowg+ard Blue Plus Multi-Lumen Central Venous Catheterization Kit. Permits venous access to central circulation by way of subclavian, jugular and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (>30 days). Clinical effectiveness of the ARROW g+ard Blue Plus catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied

Recall number
Z-1879-2017
Recalling firm
Arrow International Inc
Product
Pressure Injectable Arrowg+ard Blue Plus Multi-Lumen Central Venous Catheterization Kit. Permits venous access to central circulation by way of subclavian, jugular and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (>30 days). Clinical effectiveness of the ARROW g+ard Blue Plus catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied
Status
Terminated
Initiated
2017-03-17
Posted
2017-04-10
Terminated
2020-03-24
Reason
Arrow International is recalling due to possible voids in the catheter hub, which could lead to an interlumen crossover.
Root cause
Process control
Action
Arrow International mailed an Urgent Medical Device Notification Letter dated April 4, 2017, to affected customers to inform them of the issue. Customers were asked to discontinue use immediately and return product by completing the enclosed Recall Acknowledgement Form then emailing it to recalls@teleflex.com or faxing it to 1-855-419-8507. A Customer Service Representative will provide a Return Goods Authorization Number for the products' return. For questions regarding this recall call 866-246-6990.
Quantity
310 units
Distribution
US Distribution to NJ, IL, and PA
Lot, serial or model codes
Product Codes ASK-45703-PRWJ1, ASK-45703-PUAB1, and ASK-45703-PUPM  Lot # 13F16L0105, 23F16K0180 and 13F16L0199
510(k) numbers
["K071538"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register