FDA Medical Device Recalls (openFDA)
Ulrich Medical USA Inc: uNion Cervical Plate System Product Usage: The uNion Cervical Plate System is intended for anterior fixation of the cervical spine (C2 to Ti). The system is to be used to provide stabilization of the anterior cervical spine as an adjunct to fusion for the treatment of degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), tumors, spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), pseudarthrosis or failed previous fusion.
- Recall number
- Z-1878-2017
- Recalling firm
- Ulrich Medical USA Inc
- Product
- uNion Cervical Plate System Product Usage: The uNion Cervical Plate System is intended for anterior fixation of the cervical spine (C2 to Ti). The system is to be used to provide stabilization of the anterior cervical spine as an adjunct to fusion for the treatment of degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), tumors, spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), pseudarthrosis or failed previous fusion.
- Status
- Terminated
- Initiated
- 2017-03-07
- Posted
- 2017-04-26
- Terminated
- 2017-11-30
- Reason
- Surgical technique was revised to prevent static plate screws blocking mechanisms from disassociating.
- Root cause
- Use error
- Action
- On March 7, 2017, the recalling firm sent letters by mail informing customers of the field correction along with the updated surgical technique which contains the enhanced cautionary statement. Customers were instructed to destroy older versions of the surgical technique. The letters also contained a link to the recalling firm's website on which the recalling firm posted the current, updated copy of the surgical technique.
- Quantity
- 4,854 plates
- Distribution
- US Nationwide in the states of: AL, AR, AZ, CA, CT, FL, GA, ID, IL, KS, LA, MO, MS, NY, OH, OK, OR, TN, TX, and WI
- Lot, serial or model codes
- 043605, U010751, U010759, U011072, U012021, U012022, U012885, 043606, U010752, U010760, U012023, U012886, 043607, U010753, U010761, U012024, U012887, 043608, U010754, U010762, U012025, U012888, 043609, U010755, U010763, U012026, U012889, 043610, U010756, U010764, U012027, U012890, 043611, U010757, U010765, U012028, U012891, 043612, U010758, U010766, U012029, U012892, 043613, U010149, U010769, U012038, U012541, U012893, 043614, U010148, U010770, U012039, U012894, 043615, U010147, U010771, U012040, U012895, 043616, U010146, U010772, U012041, U012539, U012540, U012896, 043617, U010145, U010773, U012042, U012897, 043618, U010144, U010774, U010963, U012043, U012898, 043619, U010143, U010775, U012044, U012899, 043498, U010142, U012045, U012900, U101776, 043621, U010141, U010777, U012046, U012901, 043622, U010140, U010778, U012047, U012902, 043623, U010193, U010711, U010973, U011066, U012058, U012903, 043624, U010192, U010712, U010972, U011067, U012059, U012650, U012651, U012904, 043625, U010191, U010713, U010971, U011051, U012060, U012905, 043626, U010190, U010714, U010970, U012816, U012906, 043627, U010189, U010715, U012817, U012907, 043628, U010188, U010716, U010969, U012818, U012908, 043629, U010187, U010717, U010968, U012819, U012909, 043630, U010186, U010718, U010967, U012820, U012910, 043631, U010185, U010719, U010966, U012821, U012911, 043632, U010184, U010720, U012822, U012912, 043633, U010183, U010721, U010964, U012823, U012913, 043634, U010206, U011068, U012824, 043635, U010205, U011069, U012825, 043636, U010204, U011070, U012826, 043637, U010203, U011071, U012083, U012827, 043638, U010202, U012084, U012828, 043639, U010201, U011073, U012829, 043640, U010200, U011074, and U012830
- 510(k) numbers
- ["K150666"]
- PMA numbers
- not on file
- Product code
- KWQ
- Firm FEI
- 3005532635
- Firm city
- Chesterfield
- Firm state
- MO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28