FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: PANTA Nails, 10mm x 150 length, Newdeal, Lyon France; Integra, 311 Enterprise Drive, Plainsboro, NJ. The Newdeal Ankle Nail is intended for use in tibiotalocalcaneal arthrodesis and the treatment of trauma to the hindfoot and distal tibia. The nail affords rigid, load sharing fixation that incorporates a simple nail-mounted, in-line method of compression across the arthrodesis site.
- Recall number
- Z-1877-2008
- Recalling firm
- Integra LifeSciences Corp.
- Product
- PANTA Nails, 10mm x 150 length, Newdeal, Lyon France; Integra, 311 Enterprise Drive, Plainsboro, NJ. The Newdeal Ankle Nail is intended for use in tibiotalocalcaneal arthrodesis and the treatment of trauma to the hindfoot and distal tibia. The nail affords rigid, load sharing fixation that incorporates a simple nail-mounted, in-line method of compression across the arthrodesis site.
- Status
- Terminated
- Initiated
- 2008-03-24
- Posted
- 2008-09-16
- Terminated
- 2008-09-16
- Reason
- Certain PANTA Nails from one lot were incorrectly manufactured. The internal thread for certain PANTA Nails (Part Number 500050ND; Lot Number: E7XX) may present an insufficient depth and may not allow the engagement of the threaded part of the compression device. The risks presented include: an increase in surgery time, or the implantation of a nail less ideally suited for the patient's anatomy
- Root cause
- Process control
- Action
- On March 24, 2008, Recall Notification Letters and Recall Acknowledgement Forms were sent to all Integra Sales Specialists to confirm products currently in their inventory and to place inventory on hold with no further shipments to customers. If you have any questions contact Judy O'Grady at (609) 936-2311.
- Quantity
- 45 units received in US, 22 were distributed.
- Distribution
- Class II Recall - Nationwide Distribution.
- Lot, serial or model codes
- Catalog number 500050ND, Lot number: E7XX
- 510(k) numbers
- ["K050882"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28