Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics, Inc.: VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.

Recall number
Z-1876-2025
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.
Status
Open, Classified
Initiated
2025-04-14
Posted
2025-05-29
Terminated
not on file
Reason
Product was incorrectly assigned GEN 88 instead of GEN 83 on product carton, foil wrapper, and cartridge (cart) label.
Root cause
Process control
Action
An URGENT PRODUCT CORRECTION NOTIFICATION was mailed to consignees. The notification to distributors instructs consignees to forward the provided notification and Customer Confirmation Receipt to all customers who were shipped affected lots and to discontinue distributing affected lots. Affected lots in possession of distributors should be discarded. Distributors are asked to return a completed Distributor Confirmation Receipt form. The notification to customers instructs consignees to check their inventory and destroy any identified product subject to this recall. Consignees are instructed to post the provided notification near all VITROS devices and within device User Documentation until the issue has been resolved. A copy of the recall notification should be forwarded if a unit was distributed outside your facility. Consignees are to complete and return the provided Confirmation of Receipt form.
Quantity
3,587 units
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CO, FL, GA, ID, IL, IN, KS, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WI, WV & WY and the countries of Canada, Chile, Colombia, France, India, Italy, Mexico & Singapore.
Lot, serial or model codes
Catalog number: 8097990; GTIN#: 10758750004355; Coating Number: 1291; Lot No. 3788-1291-0019, 3788-1291-0052, 3788-1291-0078, 3788-1291-0079, 3788-1291-0080, 3788-1291-9975.
510(k) numbers
["K953197","K152433"]
PMA numbers
not on file
Product code
DCK
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register