GE Healthcare, LLC: ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106389-003 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106389-004 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 130 cm/ 51 in 2106389-005 ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106389-006 ECG Leadwire set, 5-lead, grouped, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106390-001 ECG Leadwire set, 3-lead, grabber, AHA, 74 cm/ 29 in 2106390-002 ECG Leadwire set, 3-lead, grabber, AHA, 130 cm/ 51 in 2106390-003 ECG Leadwire set, 3-lead, grabber, IEC, 74 cm/ 29 in 2106390-004 ECG Leadwire set, 3-lead, grabber, IEC, 130 cm/ 51 in 2106391-001 ECG Leadwire set, 5-lead, grabber, AHA, 74 cm/ 29 in 2106391-002 ECG Leadwire set, 5-lead, grabber, AHA, 130 cm/ 51 in 2106391-003 ECG Leadwire set, 5-lead, grabber, IEC, 74 cm/ 29 in 2106391-004 ECG Leadwire set, 5-lead, grabber, IEC, 130 cm/ 51 in 2106391-005 ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106391-006 ECG Leadwire set, 5-lead, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106393-001 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 130 cm/ 51 in 2106393-002 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 130 cm/ 51 in 2106393-003 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 in 2106393-004 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 74 cm/ 29 in 2106397-001 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106397-002 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106397-003 ECG Leadwire set, 6-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106398-001 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106398-002 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Recall number
Z-1876-2019
Recalling firm
GE Healthcare, LLC
Product
ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106389-003 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106389-004 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 130 cm/ 51 in 2106389-005 ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106389-006 ECG Leadwire set, 5-lead, grouped, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106390-001 ECG Leadwire set, 3-lead, grabber, AHA, 74 cm/ 29 in 2106390-002 ECG Leadwire set, 3-lead, grabber, AHA, 130 cm/ 51 in 2106390-003 ECG Leadwire set, 3-lead, grabber, IEC, 74 cm/ 29 in 2106390-004 ECG Leadwire set, 3-lead, grabber, IEC, 130 cm/ 51 in 2106391-001 ECG Leadwire set, 5-lead, grabber, AHA, 74 cm/ 29 in 2106391-002 ECG Leadwire set, 5-lead, grabber, AHA, 130 cm/ 51 in 2106391-003 ECG Leadwire set, 5-lead, grabber, IEC, 74 cm/ 29 in 2106391-004 ECG Leadwire set, 5-lead, grabber, IEC, 130 cm/ 51 in 2106391-005 ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106391-006 ECG Leadwire set, 5-lead, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106393-001 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 130 cm/ 51 in 2106393-002 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 130 cm/ 51 in 2106393-003 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 in 2106393-004 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 74 cm/ 29 in 2106397-001 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106397-002 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106397-003 ECG Leadwire set, 6-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106398-001 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106398-002 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.
Status
Terminated
Initiated
2019-05-10
Posted
not on file
Terminated
2024-05-24
Reason
ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Root cause
Process design
Action
The firm, GE Healthcare, initiated a "URGENT MEDICAL DEVICE CORRECTION"letter dated 05/10/2019. The letter explained the reason for recall and provided the following safety instructions: The affected ECG trunk cables and leadwires may continue to be used for monitoring only. Discontinue use of the affected ECG trunk cables and leadwires for patients where an arrhythmia that might require defibrillation is foreseeable, and in these patients use unaffected ECG trunk cables and leadwires. If defibrillation turns out to be unexpectedly needed when the affected ECG trunk cables and leadwires are being used just for monitoring, follow the instructions below: 1) Disconnect ALL ECG leadwires from the patient. 2) Defibrillate the patient per hospital protocol. 3) If it is possible to monitor the patients rhythm using defibrillator ECG leadwires or pads please do so. If this is not possible, reconnect ECG leadwires after the patients rhythm has been defibrillated. The recalled products will be replaced. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
58,045 units (31,108 US; 26937 OUS) in total
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Lot, serial or model codes
product does not have lot or serial codes  REF/Catalog Number GTIN/UDI 2106389-001 00840682139960 2106389-002 00840682139779 2106389-003 none  2106389-004 none 2106389-005 00840682139786 2106389-006 none 2106390-001 00840682139687 2106390-002 00840682140232 2106390-003 none 2106390-004 none 2106391-001 00840682140799 2106391-002 00840682139830 2106391-003 none 2106391-004 none 2106391-005 00840682140348 2106391-006 none  2106393-001 00840682140287 2106393-002 none 2106393-003 00840682139885 2106393-004 none 2106397-001 00840682140522 2106397-002 00840682140461 2106397-003 00840682140041 2106398-001 none 2106398-002 none
510(k) numbers
not on file
PMA numbers
not on file
Product code
DSA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register