Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc. Cardiac Rhythm Disease Management: Medtronic Mullins Transseptal Adult 7F, REF 008551. Intended for the introduction of cardiovascular catheters into the left side of the heart.

Recall number
Z-1876-2013
Recalling firm
Medtronic Inc. Cardiac Rhythm Disease Management
Product
Medtronic Mullins Transseptal Adult 7F, REF 008551. Intended for the introduction of cardiovascular catheters into the left side of the heart.
Status
Terminated
Initiated
2013-07-08
Posted
2013-08-05
Terminated
2014-03-06
Reason
Medtronic identified a specific labeling issue for specific lots of Mullins Transseptal Catheter Introducer Sheath, model 008551, in which packages labeled as Mullins 7F (French) may actually contain an 8F(French) dilator. Medtronic is requesting the return of potentially mis-labeled Mullins Transseptal Catheter Introducer 7 French Sheaths. Due to the differences in size, an 8F dilator will not
Root cause
Packaging change control
Action
Consignees were sent on 7/8/2013 a Medtronic, Inc " Urgent Medical Device Recall" letter dated 8 July 2013. The letter was addressed to Healthcare Professionals. The letter identified the problem and the product involved in the recall. Advised consignees to remove and quarantine the product from their inventory and return the product. Requested consignees to complete the attached Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612. If the product has been further distributed advised consignees to please notify them of the recall.
Quantity
269
Distribution
Worldwide Distribution-USA (nationwide) including DC and the states of AZ, AK, CA, DE, GA, IL, KY, MA, MI, MN, MS, NV, NY, NC, OH, PA, TN, and TX, and the countries of AUSTRALIA, BRAZIL, CHILE, INDIA, and SINGAPORE.
Lot, serial or model codes
Lot Numbers: 50672012, 50676944, 50683123, 50693426, 50710622.
510(k) numbers
["K791963"]
PMA numbers
not on file
Product code
DYB
Firm FEI
2182208
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register