GE Healthcare, LLC: ECG Trunk Cables: REF/Catalog Number Description 2106305-001 ECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft. 2106305-002 ECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft. 2106305-003 ECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft. 2106305-004 ECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft. 2106306-001 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft. 2106306-002 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft. 2106306-003 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft. 2106306-004 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 1.2 m/4 ft. 2106307-001 ECG Trunk Cable, 6-lead, AHA, 3.6 m/12 ft. 2106307-002 ECG Trunk Cable, 6-lead, AHA, 1.2 m/4 ft. 2106307-003 ECG Trunk Cable, 6-lead, IEC, 3.6 m/12 ft. 2106307-004 ECG Trunk Cable, 6-lead, IEC, 1.2 m/4 ft. 2106308-001 ECG Trunk Cable, 12-lead, AHA, 3.6 m/12 ft. 2106308-002 ECG Trunk Cable, 12-lead, AHA, 1.2 m/4 ft. 2106308-003 ECG Trunk Cable, 12-lead, IEC, 3.6 m/12 ft. 2106308-004 ECG Trunk Cable, 12-lead, IEC, 1.2 m/4 ft. 2106310-001 ECG Trunk Cable, 3/5-lead w/ESU filter, AHA, 3.6 m/12 ft. 2106310-002 ECG Trunk Cable, 3/5-lead w/ESU filter, IEC, 3.6 m/12 ft. 2106311-001 ECG Trunk Cable, 3/5-lead, AHA, 6 m/20 ft. 2106311-002 ECG Trunk Cable, 3/5-lead, IEC, 6 m/20 ft. The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable.

Recall number
Z-1875-2019
Recalling firm
GE Healthcare, LLC
Product
ECG Trunk Cables: REF/Catalog Number Description 2106305-001 ECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft. 2106305-002 ECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft. 2106305-003 ECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft. 2106305-004 ECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft. 2106306-001 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft. 2106306-002 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft. 2106306-003 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft. 2106306-004 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 1.2 m/4 ft. 2106307-001 ECG Trunk Cable, 6-lead, AHA, 3.6 m/12 ft. 2106307-002 ECG Trunk Cable, 6-lead, AHA, 1.2 m/4 ft. 2106307-003 ECG Trunk Cable, 6-lead, IEC, 3.6 m/12 ft. 2106307-004 ECG Trunk Cable, 6-lead, IEC, 1.2 m/4 ft. 2106308-001 ECG Trunk Cable, 12-lead, AHA, 3.6 m/12 ft. 2106308-002 ECG Trunk Cable, 12-lead, AHA, 1.2 m/4 ft. 2106308-003 ECG Trunk Cable, 12-lead, IEC, 3.6 m/12 ft. 2106308-004 ECG Trunk Cable, 12-lead, IEC, 1.2 m/4 ft. 2106310-001 ECG Trunk Cable, 3/5-lead w/ESU filter, AHA, 3.6 m/12 ft. 2106310-002 ECG Trunk Cable, 3/5-lead w/ESU filter, IEC, 3.6 m/12 ft. 2106311-001 ECG Trunk Cable, 3/5-lead, AHA, 6 m/20 ft. 2106311-002 ECG Trunk Cable, 3/5-lead, IEC, 6 m/20 ft. The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable.
Status
Terminated
Initiated
2019-05-10
Posted
not on file
Terminated
2024-05-24
Reason
ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Root cause
Process design
Action
The firm, GE Healthcare, initiated a "URGENT MEDICAL DEVICE CORRECTION"letter dated 05/10/2019. The letter explained the reason for recall and provided the following safety instructions: The affected ECG trunk cables and leadwires may continue to be used for monitoring only. Discontinue use of the affected ECG trunk cables and leadwires for patients where an arrhythmia that might require defibrillation is foreseeable, and in these patients use unaffected ECG trunk cables and leadwires. If defibrillation turns out to be unexpectedly needed when the affected ECG trunk cables and leadwires are being used just for monitoring, follow the instructions below: 1) Disconnect ALL ECG leadwires from the patient. 2) Defibrillate the patient per hospital protocol. 3) If it is possible to monitor the patients rhythm using defibrillator ECG leadwires or pads please do so. If this is not possible, reconnect ECG leadwires after the patients rhythm has been defibrillated. The recalled products will be replaced. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
58,045 units (31,108 US; 26937 OUS) in total
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Lot, serial or model codes
Product does not have lot or serial numbers  REF/Catalog Number GTIN/UDI 2106305-001 00840682137775 2106305-002 00840682137898 2106305-003 none 2106305-004 none 2106306-001 00840682137843 2106306-002 00840682137737 2106306-003 none 2106306-004 none 2106307-001 00840682137836 2106307-002 00840682137751 2106307-003 none 2106307-004 none 2106308-001 00840682137713 2106308-002 00840682137812 2106308-003 none 2106308-004 none 2106310-001 00840682137935 2106310-002 none 2106311-001 00840682137676 2106311-002 none
510(k) numbers
not on file
PMA numbers
not on file
Product code
DSA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register