FDA Medical Device Recalls (openFDA)

Hamilton Medical AG: HAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.

Recall number
Z-1874-2019
Recalling firm
Hamilton Medical AG
Product
HAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.
Status
Terminated
Initiated
2019-03-25
Posted
2019-07-11
Terminated
2024-06-03
Reason
New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.
Root cause
Software design
Action
On 03/25/19, the firm, Hamilton AG, issued Medical Device Field Safety Corrective Action notices to customers via certified mail with receipt response. Customers were advised that if ventilators entered the ambient state, they should provide alternate ventilation immediately. In addition, in order to exit the ambient state, you must turn off ventilator power. The firm has developed a new software version, which reduces the probability of the ventilator entering an ambient state due to the occurrence of an error which leads to the user message "panel connection lost". The firm will contact customers to schedule the software update. The firm requests that consignees distribute this notification to all end users within their organization. They also request that consignees complete and return the acknowledgement form via email: nicole.robinson@hamiltonmedical.net; fax: (775) 856-5621, or mail: Hamilton Medical, Inc., PO Box 30008, Reno, NV 89520. Customers with questions are encouraged to call 1-800-426-6331, option #2. In addition, with the Medical Device Field Safety Corrective Action notices, the firm also provided customers with Medical Device Field Safety Corrective Action notices, dated 02/11/19, sent to OUS customers for Hamilton-G5 and Hamilton-S1 ventilators. Customers were advised that the Swiss firm has developed a new software for the HAMILTON-G5/S1 ventilators. The software version 2.80 reduces the probability of the ventilator entering an ambient state due to the occurrence of an error which leads to the user message "panel connection lost". When the HAMILTON-G5/S1 ventilator enters an ambient state, the following conditions apply: - The HAMILTON-G5/S1 releases both a visual and an acoustical high-priority alarm to alert medical staff. - The inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. - Provide alternative ventilation immediately. - You must turn off ventilator power to exit the Am
Quantity
4338 units
Distribution
Worldwide Distribution- US (Nationwide Distribution): MI, WI, OR, MO, TX, NV, PA, SC, HI, MN, ND, NC, AZ, FL, CA, MA, GA, NY, VA, WA, IN, CO, LA, IL, OH, TN, SD, OK, MI, AR, C, AK, WV, NJ, IA, NE, KS, MT, NH, AL, MD, KY, CT, and RI. And (International Distribution): AT, BE, BG, CH, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HU, IT, MT, NL. NO, PL.PT, RO, SE, SI, SK, TR, HR. AGO, ARE, ARM, AUS, AZE, BGD, BHR , BLR, BOL, BRA, CAN, CHL, CHN, COL, CRI, DOM, DZA, ECU, EGY, EST, HKG, IDN, IND, IRQ, ISR, JOR, JPN, KAZ, KOR, KWT, LBN, LBY, LKA, LVA, MAR, MDA, MDV, MEX, MNG, MYS, NPL, NZL, OMN, PAK, PAN, PER, PHL, PRI, PSE, QAT, RUS, SAU, SGP, SUR, SYR, THA, TUN, TWN, UKR, URY, USA, VEN, VNM, YEM, and ZAF.
Lot, serial or model codes
Serial numbers 16192 and lower
510(k) numbers
["K081521","K103803","K131774"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3001421318
Firm city
Bonaduz
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register