Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: Integra Cranial access kit (without Prep solutions) Bit and Guard- Used during a ventriculostomy procedure. Catalog Number: INS5HND

Recall number
Z-1873-2024
Recalling firm
Integra LifeSciences Corp.
Product
Integra Cranial access kit (without Prep solutions) Bit and Guard- Used during a ventriculostomy procedure. Catalog Number: INS5HND
Status
Open, Classified
Initiated
2024-04-17
Posted
2024-05-20
Terminated
not on file
Reason
Potential holes and tears in the sterile packaging (header bag) of the Cranial Access Kit may compromise the sterility of the device.
Root cause
Process control
Action
Integra LifeSciences issued First Notification: Urgent: Voluntary Medical Device Recall Letter delivered by courier service, facsimile or email between April 17th and April 18th, 2024. Letter states reason for recall, health risk and action to take: 1. If you do have the affected product listed in Table 1, remove the product from further distribution. 2. Complete the attached Acknowledgment Form. 3. If you do have affected product, check the box I do have affected product. Record the total quantity of affected product you have. 4. If you do not have affected product, check the box, I do not have affected product. 5. After completion, please email the Acknowledgement Form to FCA3@integralife.com or FAX to 1-609-750- 4220. 6. Keep a copy of the form for your records. 7. Please check your customer traceability records for shipments of above catalog and lot numbers. 8. Forward a copy of the enclosed Field Safety Notice to any of your customers that have purchased the above catalog and lot number. 9. When the form from you and your customer is received, and it is noted that you or your customer has affected product, Customer Service will contact you or your customer and provide an RMA number and directions to return the product. You can request a credit for the quantities returned. Receipt of this form ensures that Integra has achieved a level of effectiveness in communicating this information. We recommend you also maintain a copy of this notification and the signed copy of the acknowledgement form for your records. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com
Quantity
1,822 units
Distribution
Worldwide distribution - US Nationwide and the countries of CA, BE, CL.
Lot, serial or model codes
UDI-DI: 10381780263906 Lot Number/Exp. Date: 7257032 07/16/2025 ; 7275713 09/03/2025 ; 7291974 09/03/2025; 7312131 09/03/2025; 7289690 09/03/2025; 7291975 09/03/2025; 7312137 09/03/2025 7312139 09/03/2025
510(k) numbers
["K961113"]
PMA numbers
not on file
Product code
HBG
Firm FEI
3003418325
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register