Skip to the content

FDA Medical Device Recalls (openFDA)

Cardinal Health 200, LLC: Catalog No. 8888160325; Umbilical Vessel Catheter 2.5 French single lumen

Recall number
Z-1871-2025
Recalling firm
Cardinal Health 200, LLC
Product
Catalog No. 8888160325; Umbilical Vessel Catheter 2.5 French single lumen
Status
Open, Classified
Initiated
2025-05-02
Posted
2025-05-29
Terminated
not on file
Reason
Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.
Root cause
Package design/selection
Action
The firm started notifying customers on May 2, 2025 via Urgent Medical Device Product Recall letters. Customers were instructed to remove any devices currently in use. In addition, customers should review their inventory and segregate/quarantine all affected product. If product has been further distributed, the distributor should notify all downstream customers. Customers are asked to return an acknowledgement form whether they have affected product or not. Customers should contact the firm to arrange for return and credit or questions regarding suitable alternative products. For questions related to this recall, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Quantity
1710
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Brazil, Canada, Chile, Colombia, Singapore, South Africa.
Lot, serial or model codes
UDI-DI (ea) 10192253040296; UDI-DI (box) 50192253040294 Lots 2433100060 2433100059 2422700145 2422700139 2422700146 2422700142
510(k) numbers
["K130725"]
PMA numbers
not on file
Product code
FOS
Firm FEI
3001236905
Firm city
Waukegan
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register