Philips North America Llc: Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143

Recall number
Z-1871-2024
Recalling firm
Philips North America Llc
Product
Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143
Status
Open, Classified
Initiated
2024-04-15
Posted
2024-05-20
Terminated
not on file
Reason
Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity
Root cause
Under Investigation by firm
Action
Philips issued URGENT Medical Device Correction Letter on 4/15/24 to US consignees. [International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements]. Letter states reason for recall, health risk and action to take: " Please continue to use your system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Place this URGENT Medical Device Correction Letter with your system documentation. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken . Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a solution to address the issue (reference FCO72800819 and FCO72800820). If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Quantity
18 units
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
UDI-DI: (01) 00884838085015 (21) + Serial number Serial Numbers- Applied to FCO72800819: 500046, 500064, 500074, 500102, 500110, 500111, 500145, 500168, 500178, 500211 500216, 500217, 500222, 500225, 500230, 500240, 500242. Serial Number-Applied to FCO72800820 : 500615
510(k) numbers
["K180015"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register